Class I FDA device recall · Z-0003-2025

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

MiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G, REF: MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892; MiniMed 780G, REF: MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896

Why it was recalled

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency with which the user must insert a new battery. Further, the defect may also deliver the low battery alarm when there is significantly less time before the battery completely runs out, from the expected up to 10-hour buffer the low alert is supposed to signal, leading to device powering down and stopping insulin delivery unexpectedly. Per the firm, a subset of pumps with the defect showed the low alert alarm was delivered ~2.5 hours before the device shut off, however, patient reports indicate the time could be even shorter. Replacing the battery will not resolve this issue and the short battery life and the truncated low battery alert lead time will continue to occur. Early battery depletion could result in power loss and insulin delivery could be interrupted, resulting in under delivery of insulin potentially leading to hyperglycemia and/or DKA.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
REF/UDI-DI(GTIN):  MMT-1780/00763000190453, 00643169946057, 00763000365882, 00763000096984, 00763000283445, 763000283513, 763000283506, 763000365882, 763000283520, 76300028351303, 763000393786, 763000166526, 76300028351301, 643169939202, 763000190460, 763000090203, 763000256821, 643169939219, 763000072537, 763000365912, 00763000090203, 00763000190460, 00763000283520, 00763000393786;  MMT-1781/00763000187415, 00763000283568, 00763000283544, 00763000404536, 00763000283551, 00763000283537, 00763000365899, 00763000371937, 00763000225575, 00763000225582, 00763000235765, 00763000187378, 00763000140601, 763000240707, 763000187378, 763000140601, 763000283544;  MMT-1782/00763000187422, 00763000235772, 00763000283599, 00763000264727, 00763000404543, 00763000283582, 00763000371944, 00763000365905, 00763000283575, 00643169957640, 00763000157463, 00763000225612, 763000225629, 763000225612, 763000264727;  MMT-1741/00763000408916, 00763000408930;  MMT-1742/00763000408947, 00763000250232, 00763000408954;  MMT-1760/763000278427, 763000283445, 643169946057, 763000190453, 763000096984, 643169946040, 00643169946057, 00763000190453, 00763000283445, 763000367084, 763000179649, 763000395858, 00763000096984;  MMT-1761/763000187392, 00763000165451, 00763000165420, 00763000165468, 00763000165444, 00763000165437, 00763000257309, 00763000165413, 00763000289065, 00763000289089, 00763000289041, 00763000367657, 00763000289195, 00763000289096, 00763000289072, 00763000289058, 00763000324797, 00763000374518, 00763000374532, 00763000374525, 00763000374556, 00763000399849, 00763000399856, 00763000374587, 763000367657;  MMT-1762/763000264710, 00763000264710, 00763000165475, 00763000165482, 00763000165499, 00763000165505, 00763000165512, 00763000165529, 00763000165536, 00763000165543, 00763000165550, 00763000203313, 00763000216481, 00763000252076, 00763000283476, 00763000289102, 00763000289119, 00763000289126, 00763000289133, 00763000289140, 00763000289164, 00763000289171, 00763000289188, 00763000324780, 00763000374594, 00763000374600, 00763000374624, 00763000399986;  MMT-1880/00763000414344, 00763000545604, 00763000439859, 763000578640, 763000586195, 76300058619502, 763000439866, 763000413750, 76300057864002, 763000414344, 763000545604, 763000640095, 763000439859, 76300058619501, 00763000413750, 00763000586188;  MMT-1881/00763000416546, 00763000439903, 00763000629892, 763000416546;  MMT-1882/00763000416553, 00763000439910, 00763000545505, 00763000629908;  MMT-1890/763000545604, 763000414344, 763000439859, 763000640095, 00763000414344, 00763000439859, 00763000545604, 76300054560402, 00763000640095, 76300064009502;  MMT-1891/00763000642242, 00763000456634, 00763000642235, 00763000418151, 00763000482596;  MMT-1892/00763000416553, 00763000439910, 00763000545505, 00763000629908;  MMT-1884/00763000414344, 76300081937801, 76300081938501, 763000819378, 763000819385, 00763000586195, A7630007585401, 00763000758547, 00763000439866, 00763000578640, 00763000439859, 00763000545604, 00763000640095, 00763000819392, 763000545604, 00763000758530, 00763000504762;  MMT-1885/00763000384258, 00763000521523, 00763000442187, 00763000521509, 00763000578671, 00763000639228, 00763000421939, 00763000597900, 00763000854942, 00763000854829, A7630006392201, 763000384272, 763000521523, 763000854829;  MMT-1886/00763000384265, 00763000441449, 00763000521530, 00763000521516, 00763000578688, 00763000442194, 00763000639235, 763000384289, 763000521530, 76300044240801, 763000688165, 76300063923501;  MMT-1894/76300086566501, 76300075854701, 763000758547, 763000819392, 763000865665, 00763000758547, 00763000819392;  MMT-1895/00763000411411, 00763000734435, 00763000411473, 00763000411466, 00763000541569, 00763000411480, 00763000420703, 00763000411428, 00763000596217, 00763000411404, 00763000533861, 00763000411381, 00763000411503, 00763000596187, 00763000411398, 00763000596194, 00763000411374, 00763000411442, 00763000411435, 00763000411497, 00763000411459, 00763000596224, 00763000411633, 00763000596231, 00763000639266, 00763000384272, 00763000521547, 00763000732585, 763000734435;  MMT-1896/763000521554,00763000411510, 00763000411565, 00763000411640, 00763000411541, 00763000411589, 00763000734442, 00763000411527, 00763000411626, 00763000411602, 00763000411558, 00763000523305, 00763000411572, 00763000411596, 00763000914127, 00763000411619, 00763000411534, 00763000734459, 00763000544317, 00763000913250, 00763000533878, 00763000521554, 00763000384289, 00763000639273, 763000734442;  All serial numbers.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US: CT, MI, PA, WA, IA, NY, ND, AZ, TX, OH, NC, AL, MN, IN, NJ, KY, UT, CA, FL, VA, MS, NM, NV, TN, GA, MA, NH, OK, VT, IL, ME, SC, LA, WY, RI, SD, KS, WI, MD, CO, DE, AR, AK, ID, MO, NE, WV, MT, OR, DC, HI, VI, PR. OUS: Worldwide

Recall record

Recalling firm
Medtronic MiniMed, Inc.
Product type
Devices
Recall number
Z-0003-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Northridge, CA
Recall initiated
2024-07-31

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0003-2025 · Source: openFDA enforcement.