  "@context": {
    "schema": "https://schema.org/",
    "au": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native#",
    "@vocab": "https://schema.org/",
    "claims": {
      "@id": "au:claims",
      "@container": "@list"
    },
    "epistemicStatus": "au:epistemicStatus",
    "occurred": "au:occurred",
    "supported": "au:supported",
    "adjudged": "au:adjudged",
    "operative": "au:operative",
    "grounding": "au:grounding",
    "groundingType": "au:groundingType",
    "adjudicator": "au:adjudicator",
    "source": "au:source",
    "sourceUrl": "au:sourceUrl",
    "value": "au:value",
    "claimType": "au:claimType",
    "reviewStatus": "au:reviewStatus",
    "commercialInfluence": "au:commercialInfluence",
    "falsifier": "au:falsifier"
  },
  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0037-2025.jsonld",
  "@type": "schema:Dataset",
  "name": "The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.  During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.",
  "description": "FDA recall Z-0037-2025 — QUANTUM SURGICAL SAS is recalling The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.  During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data. (Devices, Class II). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0037-2025&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.  During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "QUANTUM SURGICAL SAS"
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0037-2025/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-0037-2025: QUANTUM SURGICAL SAS is recalling The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.  During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.",
      "text": "The FDA logged enforcement record Z-0037-2025: QUANTUM SURGICAL SAS is voluntarily recalling The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.  During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data. — reason: Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.. Status: Completed — the FDA has determined all recalling-firm actions are complete (the FDA's review may still be open).. Initiated 2024-08-05. Affected lots/codes: UDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0037-2025&lang=eng",
        "value": "Devices recall, status Completed",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0037-2025&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0037-2025/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-0037-2025 as Class II",
      "text": "The FDA assigned this recall Class II: Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0037-2025&lang=eng",
        "value": "Class II",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0037-2025&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0037-2025/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-0037-2025",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-0037-2025",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class II",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Completed — the FDA has determined all recalling-firm actions are complete (the FDA's review may still be open).",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "QUANTUM SURGICAL SAS",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2024-08-05",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "UDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "US Nationwide distribution in the state of FL.",
            "Status": "recorded"
          }
        }
      ]
    }
  ]
}