  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0107-2024.jsonld",
  "@type": "schema:Dataset",
  "name": "DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic",
  "description": "FDA recall Z-0107-2024 — Beckman Coulter, Inc. is recalling DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic (Devices, Class II). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0107-2024&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "Beckman Coulter, Inc."
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0107-2024/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-0107-2024: Beckman Coulter, Inc. is recalling DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic",
      "text": "The FDA logged enforcement record Z-0107-2024: Beckman Coulter, Inc. is voluntarily recalling DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic — reason: The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with  the results of sample B.. Status: Ongoing — the recall is still being carried out.. Initiated 2023-09-12. Affected lots/codes: UDI/DI 15099590732103, All DxI 9000 Access Immunoassay Analyzer running software version 1.15 or prior. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0107-2024&lang=eng",
        "value": "Devices recall, status Ongoing",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0107-2024&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0107-2024/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-0107-2024 as Class II",
      "text": "The FDA assigned this recall Class II: Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0107-2024&lang=eng",
        "value": "Class II",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-0107-2024&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-0107-2024/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-0107-2024",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-0107-2024",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class II",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Ongoing — the recall is still being carried out.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "Beckman Coulter, Inc.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with  the results of sample B.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2023-09-12",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "UDI/DI 15099590732103, All DxI 9000 Access Immunoassay Analyzer running software version 1.15 or prior",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "Worldwide - US Nationwide distribution in the state of Florida, and the countries of Croatia, Czechia, France, Germany, Ireland, Italy, Netherlands, New Zealand, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.",
            "Status": "recorded"
          }
        }
      ]
    }
  ]
}