Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
8713052U, Infusomat Large Volume Pump, Non-Wireless BATTERY PACK; Volumetric Infusion Pump System
Why it was recalled
Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
UDI-DI: 04046964708626 Serial Numbers: 878498 -881897 ***Update 8/7/24*** Additional SNs 133238, 133975, 490442, 546815, 559325, 561645, 562713, 593049, 617405, 617409, 618522, 783680, 810031
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Domestic distribution nationwide. Foreign distribution to Canada.
Recall record
- Recalling firm
- B. Braun Medical, Inc.
- Product type
- Devices
- Recall number
- Z-0124-2024
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Allentown, PA
- Recall initiated
- 2023-09-21
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-0124-2024 · Source: openFDA enforcement.