Drug & Device Safety Atlas
Class II FDA device recall · Z-0174-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Medical convenience kits Item Number/Description BUCY78E CYSTO PACK CMCP10X CYSTO PACK - 205972 IHCY10Y CYSTO PACK MHCY35AE CYSTO PACK- 212487 PLBY52E CYSTO PACK UDCY31Z CYSTO PACK - 206047 UICY28L MOR CYSTO PACK

Why it was recalled

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
BUCY78E UDI-DI 191072235350 Lot 8531411 Bag serial number 17705421;  CMCP10X UDI-DI 191072228215 Lot 8392411 Bag serial number  17622370;  IHCY10Y UDI-DI 191072159922 Lot 8026111 Bag serial number 17499185 Lot 8346211 Bag serial number 17692649 Lot 8444211 Bag serial number 17681124;  MHCY35AE UDI-DI 191072228864 Lot 7990511 Bag serial number N/A;  PLBY52E UDI-DI 191072141415 Lot 8016811 Bag serial number 17435184;  UDCY31Z UDI-DI 00191072213990 Lot 8043211 Bag serial number 17483121 17483122 17483143 17483144 17485283;  UICY28L UDI-DI 191072205537 Lot 8217311 Bag serial number  17598793  * If Bag serial number is listed as "N/A", then whole lot is affected

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Recall record

Recalling firm
American Contract Systems Inc.
Product type
Devices
Recall number
Z-0174-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Grimes, IA
Recall initiated
2025-09-04

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0174-2026 · Source: openFDA enforcement.