Drug & Device Safety Atlas
Class II FDA device recall · Z-0200-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

RX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)

Why it was recalled

RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
GTIN: 05055273206104 Lot/Batch Number: (1) 7201-0747 7201-0423 7201-0417 7201-0353 7255-0198 7255-0172 7255-0171 7255-0129 7201-1085 7201-1086 7201-1079 7201-1080 7201-0982 7201-0977 7201-0970 7201-0950 7201-0924 7201-0919 7201-0901 7201-0849 7201-0827 7201-0801 7201-0768 7201-0606 7201-0562 7201-0535 7201-0531 7201-0482  (2)¿¿¿¿¿¿¿¿¿¿¿¿¿  7201-0776 7201-0367 7201-0343¿¿¿¿¿¿¿¿¿¿¿¿¿ ¿¿¿¿ 7201-0348 ¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿¿

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution including Puerto Rico

Recall record

Recalling firm
Randox Laboratories Ltd.
Product type
Devices
Recall number
Z-0200-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Crumlin (North), N/A
Recall initiated
2023-09-06

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0200-2024 · Source: openFDA enforcement.