Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm SKUs not distributed in US
Why it was recalled
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Product Code/Product Name/UDI-DI: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 10192253006940 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 10192253006957 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 10192253006964 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 10192253006971 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 10192253007053 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 10192253006568 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 10192253007060 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 10192253006988 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 10192253006995 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 10192253007008 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 10192253007015 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 10192253012408 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 10192253012415 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 10192253012422 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 10192253012439 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm 10192253012446 All Lots
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
Recall record
- Recalling firm
- Cardinal Health 200, LLC
- Product type
- Devices
- Recall number
- Z-0211-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Waukegan, IL
- Recall initiated
- 2025-09-18
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-0211-2026 · Source: openFDA enforcement.