Class II FDA device recall · Z-0219-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.

Why it was recalled

During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI :(01)00850355006000  Lot Numbers: 2231 2231-10 2303M-11 2303M-12 2303M-13 2303M-15 2305M-01 2305M-02 2305M-03 2305M-04 2305M-05 2306M-01 2306M-02 2306M-03 2306M-04 2306M-05   2306M-06   2307M   2307M-04   2307M-05   2307M-06   2307M-07   2307M-10   2307M-11   2308M-01   2308M-02   2308M-03   2308M-07   2309M   2310M-01   2310M-03 2310M-04 2310M-05 2310M-06 2310M-08 2311M-02 2311M-03 2311M-05 2311M-09 2311M-10 2312M 2312M-01 2312M-04 2312M-05 2312M-07 2313M  2313M-01  2313M-05  2313M-07  2314M-06  2314M-07  2314M-08  2314M-09  2314M-12  2314M-13  2315M-02  2315M-03  2315M-04  2315M-05  2315M-06  2315M-07   2315M-08   2315M-13   2316M   2316M-03   2316M-04   2316M-05   2316M-06   2316M-11   2316M-12   2317M   2317M-01   2318M   2320M-06   7308M

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution.

Recall record

Recalling firm
Magellan Diagnostics, Inc.
Product type
Devices
Recall number
Z-0219-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Origin
North Billerica, MA
Recall initiated
2023-09-19

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0219-2024 · Source: openFDA enforcement.