Drug & Device Safety Atlas
Class II FDA device recall · Z-0315-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

intelliPATH Universal HRP Detection Kit, REF: IPK5011G80, containing DAB Buffer, Catalog Number: IPBF5009G20; ONCORE Pro DAB Chromogen, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60; IP DAB Chromogen Kit 80 ml, REF: IPK5010G80, containing DAB Buffer, Catalog Number: IPBF5009G20; DAB kit, REF: OPRI6056KT180, containing DAB Buffer, Catalog Number: OPRI6055T60 Folate Receptor alpha IHC Assay Kit, REF: IPI4006KG10, containing DAB Buffer, Catalog Number: IPBF5009G20

Why it was recalled

Kits containing the DAB chromogen buffer, intended for use in either manual or automated Immunohistochemistry (IHC) staining protocols, may produce weak to no staining, which may lead to test failure requiring the test to be repeated.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
REF/UDI-DI/Kit Lot(DAB Lot): IPK5011G80/00847627009005/042324(012424A),060523(051823A), 082923(062923A), 011624(072723A), 042324(012424A), 080524(032624A);  OPRI6056KT180/00847627024985/020624(111623A),032224(111623A-3);  IPK5010G80/00847627008992/071923(051823A), 091923(081023A), 112823(101923A), 100424(011524A), 041624(022624A);  OPRI6056KT180/00847627024985/071924(061824A);  IPI4006KG10/00847627008985/080323(062923A), 102723(081023A)

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution in the states of MA, AZ, CT, NY, FL, NC, IN, IL, MO, CO, TX, CA, MI, PR, PA, MD, AL, OH, KY.

Recall record

Recalling firm
Biocare Medical, LLC
Product type
Devices
Recall number
Z-0315-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Pacheco, CA
Recall initiated
2024-09-25

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0315-2025 · Source: openFDA enforcement.