Drug & Device Safety Atlas
Class I FDA device recall · Z-0531-2026

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

SJM Pericardial Patch, with EnCap Technology 9x14cm, Model Number REF C0914; PATCH, PLEDGET AND INTRACARDIAC, PETP, PTFE, POLYPROPYLENE

Why it was recalled

The impacted lot may not meet the required tensile strength specification.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI/DI 05414734007276, Lot Numbers:   T2502257, exp. 20280128; T2505144, exp. 2028.04.11

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US

Recall record

Recalling firm
Glycar SA Pty., Ltd.
Product type
Devices
Recall number
Z-0531-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
E-Mail
Origin
Irene, N/A
Recall initiated
2025-10-09

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0531-2026 · Source: openFDA enforcement.