Drug & Device Safety Atlas
Class I FDA device recall · Z-0642-2024

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

Why it was recalled

A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI 00193982465155 Serial Numbers  FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.

Recall record

Recalling firm
Globus Medical, Inc.
Product type
Devices
Recall number
Z-0642-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Audubon, PA
Recall initiated
2023-12-07

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0642-2024 · Source: openFDA enforcement.