Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.
Why it was recalled
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
1) 5555000-27: UDI/DI 840682122351, Serial Numbers: 228867XR240, 252847PP241, 281274OPT1, 281338CL240, 281VHVXR240, 305682DIG1, 305682DIG3, 305KEN240B, 401OPTI2, 405949OPT1, 408559XRDR7, 413447OPT1, 419251OPT, 419690DP1, 504842XR242, 602344240A, 602344MARY240, 713440DP2, 713440DP3, 720854DF1, 732222DP3, 763520XR10, 763520XRA6, 806725OPT, 865305DP1, 941276XRA, 970495OPTI1, 973926OPT1, BOLDWPACPORT02, GLEDWPACPORT02, HMHWT240P5, LIHOPTIMA240B, OCHSNERWB240, SAMC240A; 2) 5555000-31: UDI/DI 840682122351, Serial Numbers: 00158MOB17, DF2401800221WK, U416249MOB240; 3) 5555000-33: UDI/DI 840682122351, Serial Numbers: 082407100689, 228867XR240C; 4) 5555000-41: UDI/DI 840682122351, Serial Numbers: 00130MOB16, 5973XR0225, DEF11034, RS4025RX05, SA1058MM50; 5) 5555000-47: UDI/DI 840682122351, Serial Numbers: 206223OPTIMA1, 206248OPT2, 206341OPT2402, 212838OPTIMA1, 219326AMX2, 256HHPORT5, 256HHPORT6, 260824XR240, 270793OPT, 270793OPT1, 304256OPTIMA240, 307324XR240, 318626OPX240FP, 336716WFP2, 336716WFP3, 404616OPT2, 405657OPT, 406345OPT1, 406653XR240, 406653XRA1, 414266XR243, 423837X240, 425228OPT240, 435713OPX240, 435789XR240, 502897M241, 504842OPTFP1, 504894XR244, 513952OPT1, 530888OPT2, 541789M3, 551996DIG1, 706527OPT2, 712662OPT240, 713440DP4, 718918XR240B, 785623OPT2, 786308XR240, 801465XR240, 805652M2, 816943XR240A, 828369OPT1, 850864FWDH240, 865977DP2, 901595XRA1, 914787XR2401, 916781OPT4, 951788XR240P2, 954723XR240B, 972596OPT3, 973877OPT1, AMX240GEN2SYSTEM17, AMX240GEN2SYSTEM9, HMHXR240P7, KAFB228OPTIMA2, LNMCDP1, LVCCER240B, MBMCPORTD; 6) A0700DL: UDI/DI 840682122351, Serial Numbers: 082407250223; 7) A1700DA: UDI/DI 840682122351, Serial Numbers: 210METM2, 256543OPT, 337261OPT, 360744OPT2402, 425899OPT, 425SNORIDGE, 509946AMX1, 573814DP8, 574231OPT2, 702852OPT, 850864FWDH2401, HMHXR240P1, OMCTOPT1; 8) To be provided: UDI/DI To be provided, Serial Numbers: 256543OPT1, 405657FP1417, 513952OPT2, RU1018RX04.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide
Recall record
- Recalling firm
- GE Medical Systems, LLC
- Product type
- Devices
- Recall number
- Z-0813-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Waukesha, WI
- Recall initiated
- 2024-11-25
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-0813-2025 · Source: openFDA enforcement.