Drug & Device Safety Atlas
Class II FDA device recall · Z-0888-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography, Computed, Emission

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
1) 5178728:  UDI/DI N/A, System ID YN1200, Serial Number:  16179; System ID YN1650, Serial Number:  17339.      2) H3000WE:   UDI/DI N/A, System ID 82426060010, Serial Number Not Available; System ID 82426080021, Serial Number Not Available; System ID 82426090035, Serial Number 18605; System ID 82426090037, Serial Number 18604; System ID 82426100057, Serial Number 19191; System ID 82426150036, Serial Number 18624; System ID 82426150038, Serial Number Not Available; System ID 82426180015, Serial Number 18611; System ID 082426010021, Serial Number 18644; System ID 082426030032, Serial Number 18627; System ID 082426050012, Serial Number 18629; System ID 082426070041, Serial Number 18639; System ID 082426090044, Serial Number 18652; System ID 082426100067, Serial Number 18648; System ID 082426100069, Serial Number 18641; System ID 082426120031, Serial Number 18656; System ID 082426120033, Serial Number 18650; System ID 082426140029, Serial Number 18628; System ID 082426140030, Serial Number 18647; System ID 082426150039, Serial Number 18651; System ID 082426150040, Serial Number 18657; System ID 082426160040, Serial Number 18442; System ID 082426160054, Serial Number 18645; System ID 082426170004, Serial Number 18649; System ID 082426190031, Serial Number 18646; System ID 082426210028, Serial Number 18655; System ID 082426230028, Serial Number 18653; System ID 082426250006, Serial Number 18637; System ID 671USNHINF, Serial Number 17799; System ID A5128757, Serial Number 17566; System ID 713794INF, Serial Number 17219; System ID 920794inf, Serial Number 16364.    3) H3000WM:  UDI/DI N/A, System ID NM16353, Serial Number 16353; System ID YN1540, Serial Number 16335; System ID YN1555, Serial Number 16315.    4) H3000YG:  UDI/DI N/A, System ID A5172908, Serial Number 16143.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide

Recall record

Recalling firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Product type
Devices
Recall number
Z-0888-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Tirat Carmel
Recall initiated
2024-12-18

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0888-2025 · Source: openFDA enforcement.