Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission
Why it was recalled
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide
Recall record
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Product type
- Devices
- Recall number
- Z-0895-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Tirat Carmel
- Recall initiated
- 2024-12-18
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-0895-2025 · Source: openFDA enforcement.