Drug & Device Safety Atlas
Class II FDA device recall · Z-0963-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) with 31 infrared (IR) light-emitting diodes (LEDs) as well as an electronics control unit for the LEDs which includes its own battery power supply. The measured LED positions can be used in one of three different modes of operation: Patient tracking, Registration using surface matching, and Automatic intra-operative mask registration.

Why it was recalled

When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Part Number: 6000-390-000; UDI-DI: 37613327097628; Lots: 6000106885, 6000106894, 6000106895, 6000106896;

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Germany, Japan, South Korea.

Recall record

Recalling firm
Howmedica Osteonics Corp.
Product type
Devices
Recall number
Z-0963-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Mahwah, NJ
Recall initiated
2025-11-12

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0963-2026 · Source: openFDA enforcement.