Drug & Device Safety Atlas
Class II FDA device recall · Z-0989-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Hosptial Electronics System (PES) REF CM3000 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

Why it was recalled

As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
MN6000 (Which interfaces with CM3000 for HES) UDI-DI Code: 05414734510035 HES Serial Number/Clinic Merlin ID: M140400470 50011924 M140901162 50012084 M150102238 50013199 M150302655 50011863 M150905419 100351 M161107929 50012353 M161207976 50012166 M170108094 50011331 M170208159 50012146 M170208161 100268 M170408244 50015053 M170508275 50000752 M171108426 50051809 M171108437 50011840 M171108442 50011887 M180208509 50011887 M180308526 100710 M180608651 50011332 M180608652 50025010 M180608653 50015252 M180608669 50015252 M180608712 50060412 M180608727 50017544 M181008882 50038595 M190208990 50040301 M190208994 50040237 M190300010 50011866 M190400008 500150 M190700005 100605 M200400018 100359 M200600002 100351 M200600007 100080 M200800004 50050337 M200900014 50016848 M201100007 50050201 M210100003 9008126 M210300009 50014260 M210300011 50054572 M210300022 50013502 M210300023 50011879 M210900025 12000635 M211000003 50017405 M220600048 50025317

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.

Recall record

Recalling firm
St. Jude Medical
Product type
Devices
Recall number
Z-0989-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Atlanta, GA
Recall initiated
2024-10-05

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-0989-2025 · Source: openFDA enforcement.