Drug & Device Safety Atlas
Class II FDA device recall · Z-1042-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal

Why it was recalled

Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
1) GTIN 08714729796527, Batch Number 37370117, exp. 3-Sept-28;  2) GTIN 08714729796756, Batch Number  37416646, exp. 9-Sept-28

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL.

Recall record

Recalling firm
Boston Scientific Corporation
Product type
Devices
Recall number
Z-1042-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Maple Grove, MN
Recall initiated
2025-12-03

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1042-2026 · Source: openFDA enforcement.