Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
The Sofia 2 Flu + SARS Antigen FIA employs immunofluorescence technology in a sandwich design that is used with Sofia 2 to detect nucleocapsid protein from influenza A, influenza B, and SARS-CoV-2. This test allows for the detection of SARS-CoV and SARS-CoV-2, but does not differentiate between the two viruses.
Why it was recalled
One lot of the Sofia 2 Flu + SARS Antigen FIA (Lot 709083) was labelled with incorrect expiration dates. The kits contain pouched cassettes labeled with an expiration date of May 31, 2025, however the correct expiration date for the pouches is September 12, 2024. The kits are labeled with an expiration date of April 30, 2025, however, the correct expiration date for the kits is September 12, 2024. The issue was discovered on November 28, 2023 by a supply chain employee who determined that the kits had expiration dates beyond the FDA authorized shelf life.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
UDI-DI: 30014613339229. Lot: 709083
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution in the states of FL, TN, IL, LA, NY, IN.
Recall record
- Recalling firm
- Quidel Corporation
- Product type
- Devices
- Recall number
- Z-1086-2024
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Carlsbad, CA
- Recall initiated
- 2024-01-03
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1086-2024 · Source: openFDA enforcement.