Class II FDA device recall · Z-1153-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

ADVIA Centaur Folate 500 test kit- For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 10325366

Why it was recalled

Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI:    (01)00630414450940(10)16276336 (17)20230320;       (01)00630414450940(10)21334336(17)20230320;       (01)00630414450940(10)22079336(17)20230320;       (01)00630414450940(10)27946338(17)20230529;      (01)00630414450940(10)41225342(17)20230630;       (01)00630414450940(10)62966344 (17)20230726;    Lot Numbers: 16276336,  21334336, 22079336, 27946338, 41225342, 62966344 and higher.   Affects all current and future lots of the ADVIA Centaur Folate assay assay until the Instructions for Use are updated.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Nationwide Foreign: Albania¿¿¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Bahrain¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Bulgaria¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Cura¿ao,¿St¿Eus Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Hungary¿¿¿¿¿¿¿¿ Iran¿¿¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Jordan¿¿¿¿¿¿¿¿¿ Kenya¿¿¿¿¿¿¿¿¿¿ Latvia¿¿¿¿¿¿¿¿¿ Lithuania¿¿¿¿¿¿ Malta¿¿¿¿¿¿¿¿¿¿ Morocco¿¿¿¿¿¿¿¿ Nepal¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ Norway¿¿¿¿¿¿¿¿¿ Oman¿¿¿¿¿¿¿¿¿¿¿ Pakistan¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Saudi¿Arabia¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ U.A.E.¿¿¿¿¿¿¿¿¿ Uganda¿¿¿¿¿¿¿¿¿ Ukraine¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vatikancity¿¿¿¿ Country AR AU BR BS CA CL CN CO CR DO EC HK IL IN JP KR MX MY NZ PE PH PY SA SG TH TW UY VN ZA

Recall record

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Devices
Recall number
Z-1153-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
East Walpole, MA
Recall initiated
2023-01-09

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.