Drug & Device Safety Atlas
Class II FDA device recall · Z-1155-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.

Why it was recalled

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US: CA, LA

Recall record

Recalling firm
Canon Medical System, USA, INC.
Product type
Devices
Recall number
Z-1155-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Tustin, CA
Recall initiated
2023-11-15

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1155-2024 · Source: openFDA enforcement.