Class II FDA device recall · Z-1206-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).

Why it was recalled

After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI: 60889978621613 (Rayco) 60889978624188 (1049)  US Serial Numbers: SP0763 50066 50082 50089 50135 50145 50143 50140D 50144  OUS serial numbers: SP0819 SP0687G2 SP0726 SP0746 SP0834 SP0837 S0001 S0170 S0698G2 1085 FS0564 S0578 S0744 SP0903 SP0600 SP1035 SP0930 SP1007 SP1037 SP0772 S0934G2 S0639G2 S1020G2 S0118 S0135 S0134 S0165 S0167 S0168 S0181 S0800G2 S0784G2 S0811G2 S0832G2 S0826G2 S0822G2 S1038G2 S0678G2 S0619G2 S0636G2 S0641G2 S0643G2 S0701G2 S0738G2 S0758G2 S0760G2 S0673G2 S0785G2 S0842G2 S0860G2 S0863G2 S0888G2 S0973G2 S1105G2 S1048G2 S1073G2 S1004G2 S1110G2 S0791G2 S0941G2  SP0900 S0625

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.

Recall record

Recalling firm
Carestream Health, Inc.
Product type
Devices
Recall number
Z-1206-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Rochester, NY
Recall initiated
2023-01-30

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.