Class II FDA device recall · Z-1228-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

VITROS XT 7600 Integrated System, Catalog No. 6844461

Why it was recalled

A software defect affecting VITROS Systems running VITROS Software Versions 3.8.1 causes QC baseline statistics to not update as expected when changed by the user. This may cause erroneous patient results to be reported.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
VITROS System Software Version 3.8.1 UDI-DI 10758750031610  Serial Numbers  US:  J76000199 J76001383 J76000207 J76001385 J76000353 J76001392 J76000456 J76001404 J76000459 J76001406 J76000460 J76001516 J76000468 J76001518 J76000495 J76001519 J76000502 J76001520 J76000561 J76001581 J76001207 J76001583 J76001230 J76001611 J76001231 J76001620 J76001232 J76001644 J76001235 J76001647 J76001239 J76001660 J76001260 J76001668 J76001263 J76001671 J76001270 J76001674 J76001274 J76001692 J76001331 J76001700 J76001351 J76001748    OUS: J76000118 J76001070 J76000253 J76001074 J76000265 J76001095 J76000274 J76001121 J76000281 J76001130 J76000370 J76001152 J76000447 J76001174 J76000475 J76001328 J76000620 J76001377 J76000855 J76001378 J76000856 J76001396 J76000898 J76001399 J76000901 J76001407 J76000973 J76001408 J76000976 J76001484 J76001001 J76001673 J76001027 J76001761 J76001035

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution in the states of AZ, CA, CT, FL, IA, ID, IL, IN, KY, LA, MO, NE, NY, OH, OR, SC, SD, TX, WA, WI, WV and the countries of Australia, Brazil, Canada, China, Colombia, India, Mexico, Singapore, United Kingdom, Italy, Spain, Russia, France, and Belgium.

Recall record

Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product type
Devices
Recall number
Z-1228-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Rochester, NY
Recall initiated
2024-01-26

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1228-2024 · Source: openFDA enforcement.