Drug & Device Safety Atlas
Class II FDA device recall · Z-1241-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Optima MR360, NMRI system

Why it was recalled

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI  System IDs  082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01)  (Update 1/28/26 - Additional System ID 82427190104)

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Domestic US distribution nationwide. International distribution worldwide.

Recall record

Recalling firm
GE Healthcare (China) Co., Ltd.
Product type
Devices
Recall number
Z-1241-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Beijing, N/A
Recall initiated
2025-01-31

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1241-2025 · Source: openFDA enforcement.