Drug & Device Safety Atlas
Class II FDA device recall · Z-1279-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Cardiosave Intra-Aortic Balloon Pump (IABP) Pneumatic Interface Module (PIM) , Model Number 0997-00-1178

Why it was recalled

There is an issue related to the Safety Disk not meeting a performance requirement that may impact Intra-Aortic Balloon catheter (IAB) displacement volume. This potential failure of an impacted Safety Disk may lead to slightly reduced IAB inflation during therapy or a reduction in the degree of augmentation provided to the patient during therapy.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Serial Numbers: MH353075E2 MH353057E2 MH353067E2 MH353077E2 MH356112G2 MH353056E2 MH356111G2 MH356119G2 MH353065E2 MH356117G2

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Distribution list not yet provided.

Recall record

Recalling firm
Datascope Corp.
Product type
Devices
Recall number
Z-1279-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Mahwah, NJ
Recall initiated
2023-02-28

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1279-2023 · Source: openFDA enforcement.