Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
GE Healthcare Revolution Apex Plus, System, X-ray, Tomography, Computed
Why it was recalled
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number 973971CTREV, UDI (01)00840682118552(11)160600(21)REVVX1600010CN, Mfg. Lot/Serial Number REVVX1600010CN; System ID Number 513584APEX1, UDI (01)00840682146616(11)230300(21)REV2A2300023CN, Mfg. Lot/Serial Number REV2A2300023CN
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide distribution.
Recall record
- Recalling firm
- GE Medical Systems, LLC
- Product type
- Devices
- Recall number
- Z-1300-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Waukesha, WI
- Recall initiated
- 2025-02-18
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1300-2025 · Source: openFDA enforcement.