Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed
Why it was recalled
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 650493REV2, UDI To be provided, Mfg. Lot/Serial Number REVVX1900036CN; System ID Number 864512APEX1, UDI (01)00195278460271(11)240500(21)REV2X2400059CN, Mfg. Lot/Serial Number REV2X2400059CN; System ID Number 832325REV, UDI (01)00195278460271(11)241000(21)REV2X2400134CN, Mfg. Lot/Serial Number REV2X2400134CN; System ID Number 281MWREVCT, UDI To be provided, Mfg. Lot/Serial Number REVVX1900008CN; System ID Number 713792ACT39, UDI To be provided, Mfg. Lot/Serial Number REVV82100006CN; System ID Number 206320REV, UDI To be provided, Mfg. Lot/Serial Number 00000441459CN5; System ID Number 206320REV2, UDI To be provided, Mfg. Lot/Serial Number REVV82200025CN; System ID Number 082421190701, UDI (01)00195278633149(11)241100(21)CBFCG2400026HM, Mfg. Lot/Serial Number CBFCG2400026HM; System ID Number DK1044CT12, UDI (01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/Serial Number REV2A2000001CN; System ID Number DK1547CT02, UDI (01)00840682146616(11)210100(21)REV2A2100002CN, Mfg. Lot/Serial Number REV2A2100002CN; System ID Number 11525316, UDI To be provided, Mfg. Lot/Serial Number REV2X2400082CN; System ID Number 259056CT01, UDI (01)00195278460271(11)230800(21)REV2X2300051CN, Mfg. Lot/Serial Number REV2X2300051CN; System ID Number 11437190, UDI To be provided, Mfg. Lot/Serial Number CBGLG2400001HM; System ID Number 0850211094, UDI (01)00195278460271(11)230800(21)REV2X2300048CN, Mfg. Lot/Serial Number REV2X2300048CN; System ID Number 12622329, UDI To be provided, Mfg. Lot/Serial Number REV2X2400165CN; System ID Number M001CT06, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number M004CT02, UDI (01)00195278460271(11)230400(21)REV2X2300005CN, Mfg. Lot/Serial Number REV2X2300005CN; System ID Number AC002CT05, UDI (01)00840682123471(11)190100(21)REVV81900003CN, Mfg. Lot/Serial Number REVV81900003CN; System ID Number R002CT12, UDI (01)00195278460271(11)230500(21)REV2X2300012CN, Mfg. Lot/Serial Number REV2X2300012CN; System ID Number AC001CT05, UDI (01)00840682118552(11)181000(21)REVVX1800101CN, Mfg. Lot/Serial Number REVVX1800101CN
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide distribution.
Recall record
- Recalling firm
- GE Medical Systems, LLC
- Product type
- Devices
- Recall number
- Z-1303-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Waukesha, WI
- Recall initiated
- 2025-02-18
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1303-2025 · Source: openFDA enforcement.