Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Numerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM; (2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM; (3) REF BJ511R, Holzbach Hyst Forceps CVD245MM; (4) REF BJ520R, Heaney Hyst Forceps Grooved 205MM; (5) REF BJ521R, Heaney Hyst Forceps Grooved 250MM; (6) REF BJ522R, Heaney Hyst Forceps Grooved 195MM; (7) REF BJ524R, Heaney-Rezek Fcpsstr Ser Lgth 210MM; (8) REF BJ527R, Wiener Hyst Forceps Grooves 240MM; (9) REF BJ530R, Heaney Hysterectomy Forceps STR 225MM; (10) REF BJ531R, Heaney Hysterectomy Forceps CVD200MM; (11) REF BJ532R, Heaney Hysterectomy Forceps CVD215MM; (12) REF BJ534R, Hysterectomy For.Gwilliamsstr.205MM; (13) REF BJ550R, Wertheim Parametrium Clamp STR 250MM; (14) REF BJ551R, Wertheim Parametrium Clamp CVD245MM; (15) REF BJ552R, Wertheim Parametrium Clamp CVD240MM; (16) REF BJ553R, Wertheim Parametrium Clamphvy-CVD220MM; (17) REF MD591, Heany Ballentine Hyst Fcps STR.215MM; and (18) REF MD592, Heany Ballentine Hyst Fcps CVD.215MM.
Why it was recalled
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
GTIN numbers: (1) REF BJ500R - 04038653139378; (2) REF BJ501R - 04038653139385; (3) REF BJ511R - 04038653139392; (4) REF BJ520R - 04038653139408; (5) REF BJ521R - 04038653139415; (6) REF BJ522R - 04038653139422; (7) REF BJ524R - 04046963364250; (8) REF BJ527R - 04038653139446; (9) REF BJ530R - 04038653139453; (10) REF BJ531R - 04038653139460; (11) REF BJ532R - 04038653139477; (12) REF BJ534R - 04038653027798; (13) REF BJ550R - 04038653139484; (14) REF BJ551R - 04038653139491; (15) REF BJ552R - 04038653139507; (16) REF BJ553R - 04038653139514; (17) REF MD591 - 04046963365592; and (18) REF MD592 - 04046963365608.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.
Recall record
- Recalling firm
- Aesculap Inc
- Product type
- Devices
- Recall number
- Z-1308-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Center Valley, PA
- Recall initiated
- 2025-01-31
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1308-2025 · Source: openFDA enforcement.