Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741
Why it was recalled
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A UDI-DI (ea) 10653160383628 UDI-DI (case) 00653160383621 Kit lots 2024121390 2025033190 Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 UDI-DI (ea) 10653160368755 UDI-DI (case) 00653160368758 Kit lot 2025041090 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A UDI-DI (ea) 10653160374978 UDI-DI (case) 00653160374971 Kit lots 2024051490 2025021290 2025051390 2025091090 Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 UDI-DI (ea) 10653160375845 UDI-DI (case) 00653160375848 Kit lots 2024060690 2024101690 2024103190 2025010890 2025040190 Medline DIALYSIS CHANGE KIT SKU EBSI1453A UDI-DI (ea) 10653160383475 UDI-DI (case) 00653160383478 Kit lots 2024050790 2024051390 2024082290 2024111990 2025013090 2025032090 2025050190 2025062790 Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 UDI-DI (ea) 10653160368922 UDI-DI (case) 00653160368925 Kit lots 2025040401 2025081401 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 UDI-DI (ea) 10653160374527 UDI-DI (case) 00653160374520 Kit lots 2024061780 2024070180 2024081280 2024111290 2025012490 2025032690 2025051390 2025061790 2025071490 2025072190
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- Medline Industries, LP
- Product type
- Devices
- Recall number
- Z-1465-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Northfield, IL
- Recall initiated
- 2026-01-08
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1465-2026 · Source: openFDA enforcement.