Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.
Why it was recalled
Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
UDI-DI B13095240 Small Lot 4318 Expires 01Apr2026 Lot 4325 Expires 01Apr2026 Lot 4368 Expires 01Dec2026
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- Checkpoint Surgical Inc
- Product type
- Devices
- Recall number
- Z-1466-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Visit
- Origin
- Cleveland, OH
- Recall initiated
- 2025-02-24
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1466-2025 · Source: openFDA enforcement.