Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845
Why it was recalled
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- Medline Industries, LP
- Product type
- Devices
- Recall number
- Z-1467-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Northfield, IL
- Recall initiated
- 2026-01-08
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1467-2026 · Source: openFDA enforcement.