Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL
Why it was recalled
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
cobas pro sample supply unit, UDI-DI 07613336158852 cobas pro SSU, UDI-DI 07613336179499 all software versions previous to 03-02
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product type
- Devices
- Recall number
- Z-1476-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Indianapolis, IN
- Recall initiated
- 2026-01-20
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1476-2026 · Source: openFDA enforcement.