Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL
Why it was recalled
There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
5400-030-008 UDI-DI: 07613154340842, Lot Numbers: 2023012515 2023012715 2023022012 2023050814 2023072632 2023102312 2023102612 2024032514 2024040212; 5400-030-008ITL UDI-DI: 07613327359473, Lot Numbers: 2022092714 2023020633 2023032433 2023050915 2023052514 2023062216 2023082814 2023101112 2024031812 2022092914 2023020814 2023032814 2023051114 2023053115 2023062032 2023090618 2023101714 2024042212 2022101332 2023020915 2023033016 2023051215 2023060914 2023082215 2023100312 2023102014 2024042312 2022121932 2023021416 2023050115 2023052214 2023061412 2023082312 2023100412 2024030414 2024061814 2022122032 2023022032 2023050515 2023052412 2023061514 2023082514 2023100912 2024030512 2024062814
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
US Nationwide distribution.
Recall record
- Recalling firm
- Stryker Corporation
- Product type
- Devices
- Recall number
- Z-1479-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Portage, MI
- Recall initiated
- 2025-02-25
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·