Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
Medline medical convenience kits labeled as: 1) NEONATAL INTUBATION TRAY, REF ACC010371C; 2) NEWBORN RESUSCITATION KIT, REF ACC010381A; 3) NEWBORN RESUSCITATION KIT, REF ACC010381B; 4) NEWBORN RESUSCITATION KIT, REF ACC010381C; 5) PEDS GREY-PINK-RED, REF ACC010395; 6) PEDS PURPLE, REF ACC010397; 7) NEONATAL INTUBATION TRAY, REF ACC010541A; 8) NEONATAL INTUBATION TRAY, REF ACC010541B; 9) NEONATAL INTUBATION TRAY, REF ACC010541C; 10) PEDIATRICS INTUBATION TRAY, REF DYNJAA252.
Why it was recalled
Specific Medline Kits were manufactured using Intubation ORAL/NASAL Endotracheal Tubes which were subsequently recalled by Smiths Medical.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
1) REF ACC010371C: UDI/DI 10193489268188 (ea) 40193489268189 (case), Lot Numbers: 20DDA921, 20EDC042, 20IDC153, 21ADB113, 21CDB435, 21HDB557, 21JDA180, 21LDB244, 22HDB092, 22IDA205, 22JDA908, 22KDA049, 22LDA351, 23EDC611; 2) REF ACC010381A: UDI/DI 10193489800944 (ea) 40193489800945 (case), Lot Numbers: 20BDB854, 20FDA703; 3) REF ACC010381B: UDI/DI 10193489457803 (ea) 401934894578049 (case), Lot Numbers: 20LDA314, 20LDA466, 21ADB162, 21BDB508; 4) REF ACC010381C: UDI/DI 10193489922127 (ea) 40193489922128 (case), Lot Numbers: 21IDA608, 21JDB271, 21KDA979, 21LDA450, 22BDA818, 22FDB822, 22KDB696, 23ADA938, 23CDA005, 23CDB015, 23EDA977, 23FDB762, 23IDB196, 23LDA440, 23LDB579, 24CDA637, 24GDB084, 24IDB423, 24JDB726, 24KDB730; 5) REF ACC010395: UDI/DI 10193489218152 (ea) 40193489218153 (case), Lot Numbers: 20BDC901; 6) REF ACC010397: UDI/DI 10193489218169 (ea) 40193489218160 (case), Lot Numbers: 20CDA453, 20CDA459, 20EDB330; 7) REF ACC010541A: UDI/DI 10195327036980 (ea) 40195327036981 (case), Lot Numbers: 22ADA311; 8) REF ACC010541B: UDI/DI 10195327153649 (ea) 40195327153640 (case), Lot Numbers: 22EDB187; 9) REF ACC010541C: UDI/DI 10198459155765 (ea) 40198459155766 (case), Lot Numbers: 24JDC035; 10) REF DYNJAA252: UDI/DI 10193489297980 (ea) 40193489297981 (case), Lot Numbers: 20EBC867, 20EBD635, 20FBJ920, 22FBF130, 22IBP405
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
California
Recall record
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product type
- Devices
- Recall number
- Z-1505-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Origin
- Northfield, IL
- Recall initiated
- 2025-02-24
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1505-2025 · Source: openFDA enforcement.