Class I FDA device recall · Z-1545-2026

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX Kit/RE4001-01, Leader TRUE METRIX Kit/RE4002-01, Leader TRUE METRIX Meter Only/RE4002-40, CVS TRUE METRIX Kit/RE4007-01SB, GNP TRUE METRIX Meter Only/RE4011-00, GNP TRUE METRIX Kit /RE4011-01, MEIJER TRUE METRIX Kit/RE4019-01, FRED'S TRUE METRIX Kit/RE4023-01, DISCOUNT DRUG TRUE METRIX Kit/RE4025-01, KINRAY TRUE METRIX Kit/RE4027-01, HEB TRUE METRIX Kit/RE4031-01, McKesson TRUE METRIX Starter Kit/RE4051-43, SUNMARK TRUE METRIX Meter Only/RE4054-00, SUNMARK TRUE METRIX Kit/RE4054-01, Signature Care TRUE METRIX Kit/RE4060-01, CareOne TRUE METRIX Kit/RE4061-01, RITE AID TRUE METRIX Kit/RE4066-01, Liberty TRUE METRIX Meter Only/RE4078-40, HUMANA TRUE METRIX Starter Kit/RE4081-03, Wegmans TRUE METRIX Kit/RE4087-01, HEALTHMART TRUE METRIX Meter Only/RE4089-00, HEALTHMART TRUE METRIX Kit/RE4089-01, ARRIVA TRUE METRIX Meter Only/RE4095-40, HealthyAccents TRUE METRIX Kit/RE4097-01, Shopko TRUE METRIX Kit/RE4098-01, Henry Schein TRUE METRIX Kit/RE4099-40, Henry Schein TRUE METRIX Meter Only/RE4099-45, Millennium Pharmay Services TRUE METRIX Kit/RE4202-43, Kroger TRUE METRIX Kit/RE4203-01, TOPCARE TRUE METRIX Kit/RE4209-01, Foster and Thrive TRUE METRIX Kit/RE4211-01, TRUE METRIX Meter Only/RE4H01-00, TRUE METRIX Kit/RE4H01-01, TRUE METRIX Starter Kit/RE4H01-04, TRUE METRIX NFRS Meter Only/RE4H01-40, TRUE METRIX NFRS Starter Kit/RE4H01-43, TRUE METRIX Kit/RE4H01-51, TRUE METRIX Meter Only (mg/dL)/RE4i01-00, TRUE METRIX Kit (mg/dL)/RE4i01-01, TRUE METRIX Kit (mg/dL)/RE4i01-02, TRUE METRIX Starter Kit (mg/dL)/RE4i03-02, TRUE METRIX Kit (mmol/L)/RE4i03-11, Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00, Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00, TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00, TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11, TRUE METRIX Kit (mg/dL)/RE4i61-02, TRUE METRIX Kit (mg/dL)/RE4i61-02THI, TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11, TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12.

Why it was recalled

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940.  International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/A, RE4i23-00/21292010133, RE4i29-11/N/A, RE4i61-02/N/A, RE4i61-02THI/N/A, RE4i82-11/21292012205, RE4i82-12/21292009977.  Owner's Booklet or IFU Description/Part and Revision: TRUE METRIX Owner's Booklet/ RE4TVH03 Rev 56 and prior, TRUE METRIX Owner's Booklet (English)/ RE4TVH35 Rev 51 and prior,  TRUE METRIX Owner's Booklet (Spanish)/ RE4TVH35S Rev 51 and prior,   McKesson Med Surg TRUE METRIX Self Monitoring Owners Booklet/ RE4SUN03 Rev 55 and prior,  McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior,  TRUE METRIX Blood Glucose Monitoring System IFU Trividia House Brand/ RE4ITV16 Rev 53 and prior,  TRUE METRIX System Instructions for Use (IFU) for Trividia Health UK Limited/RE4UKT16 Rev 55 and prior,  Trividia Health LATAM - TRUE METRIX System IFU mg/dL/ RE4LAT16 Rev 50 and prior,  TRUE METRIX System Instructions for Use (IFU) for Trividia Health Australia/ RE4AUT16 Rev 56 and prior.  Device serial numbers: Prefix "T" followed by an eight-digit numeric sequence.

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.

Recall record

Recalling firm
Trividia Health, Inc.
Product type
Devices
Recall number
Z-1545-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Fort Lauderdale, FL
Recall initiated
2026-02-06

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.