Drug & Device Safety Atlas
Class II FDA device recall · Z-1558-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

DEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR

Why it was recalled

The instruction manual is being replaced with an updated version that is in alignment with labeling requirements specified in the IEC-60101-1 standard.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Model 1025DS - UDI-DI: 00885304000846; Serial Numbers A175230001DS to A232160185DS;  Model 1025UK - UDI-DI: 00885304022480; Serial Numbers A179210001US to A232060432US;  Model 1025KS - UDI-DI: 00853040224734; Serial Numbers A176060001KS to A232220085KS;  Model 1025KS-AR - UDI-DI: 00853040224734; Serial Numbers A216240003KS to A229210010KR

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution.

Recall record

Recalling firm
DeVilbiss Healthcare LLC
Product type
Devices
Recall number
Z-1558-2023
Voluntary / mandated
Voluntary: Firm initiated
Origin
Somerset, PA
Recall initiated
2023-03-09

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1558-2023 · Source: openFDA enforcement.