Class II FDA device recall · Z-1584-2023

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Spectral CT 7500 Computed Tomography X-Ray System, Model Number 728333

Why it was recalled

Pressing the Unload Pedal of the foot switch to perform the unload function may cause entrapment of the operators foot. If foot or leg entrapment occurs while pressing Unload Pedal, the risk to operators may include: pain, muscle or ligament sprain, abrasion, laceration, contusion, fracture, loss of function, and joint dislocation.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI: 00884838101111; Serial Numbers: 10058 10074 10080 10093 10096 10097 10098 10099 10100 10102 10103 10104 10105 10106 10107 10109 10110 10111 10114 10115 10117 10119 10123 396001 396003 10014 10015 10016 10017 10018 10019 10020 10021 10022 10023 10024 10025 10027 10031 10032 10033 10034 10035 10036 10038 10039 10040 10041 10042 10043 10044 10045 10046 10047 10048 10050 10051 10052 10053 10055 10056 10057 10059 10060 10061 10062 10063 10065 10066 10067 10068 10069 10070 10075 10076 10077 10081 10082 10085 10086 10087 10089 10090 10092 10094 10095 10108 10116 10120 10049 10121 10124

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution including in the states of States: AZ, IN, MA, MD, MN, NY, OH, PA, and TX. The countries of Australia, Austria, China, Czech Republic, Denmark, France, Germany, Israel, Japan, Malaysia, Netherlands, Norway, Palestine, Panama, South Korea, Spain, Switzerland, Thailand, United Kingdom.

Recall record

Recalling firm
Philips North America Llc
Product type
Devices
Recall number
Z-1584-2023
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Cambridge, MA
Recall initiated
2023-03-17

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.