Drug & Device Safety Atlas
Class II FDA device recall · Z-1634-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Brand Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Product Name: TECNIS Eyhance IOL with TECNIS Simplicity" Delivery System Model/Catalog Number: DIB00 Software Version: N/A Product Description: The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. The lens is preloaded and pre-assembled in the delivery system. The TECNIS SIMPLICITY" Delivery System Model DIB00 contains the TECNISEyhance" IOL, which is a one-piece, foldable, posterior chamber lens with an overall diameter of 13.0 mm and an optic diameter of 6.0 mm. The TECNIS SIMPLICITY" Delivery System is designed to provide a sterile,controlled and touch-free method of delivering the lens into the eye. Component: No

Why it was recalled

Due to identifying a limited quantity of intraocular lenses that may have a haptic that sticks to the optic which prevents the lens from unfolding as consistently as expected.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
Lot Code: GTIN: 05050474731776 SKU: DIB00U0210-12 Serial Numbers: 2409342403 2410072403 2409612403 2409692403 2409752403 2891682403 2891132403 2892032403 2531882403 2531902403 2531212403 2530842403 2531242403 2531892403 2531622403

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

U.S Nationwide distribution in the states of CA,FL, IL, ME,MS, NJ, NY, OK, OR, TX, and WI.

Recall record

Recalling firm
AMO Puerto Rico Manufacturing, Inc.
Product type
Devices
Recall number
Z-1634-2026
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Anasco, PR
Recall initiated
2026-02-19

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1634-2026 · Source: openFDA enforcement.