  "@context": {
    "schema": "https://schema.org/",
    "au": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native#",
    "@vocab": "https://schema.org/",
    "claims": {
      "@id": "au:claims",
      "@container": "@list"
    },
    "epistemicStatus": "au:epistemicStatus",
    "occurred": "au:occurred",
    "supported": "au:supported",
    "adjudged": "au:adjudged",
    "operative": "au:operative",
    "grounding": "au:grounding",
    "groundingType": "au:groundingType",
    "adjudicator": "au:adjudicator",
    "source": "au:source",
    "sourceUrl": "au:sourceUrl",
    "value": "au:value",
    "claimType": "au:claimType",
    "reviewStatus": "au:reviewStatus",
    "commercialInfluence": "au:commercialInfluence",
    "falsifier": "au:falsifier"
  },
  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1662-2025.jsonld",
  "@type": "schema:Dataset",
  "name": "CADD Solis HSPCA Pump  ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:  PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA ¿ 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA ¿ 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA ¿ 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH     1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA ¿ 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA ¿ 21-2112-0100-51",
  "description": "FDA recall Z-1662-2025 — Smiths Medical ASD, Inc. is recalling CADD Solis HSPCA Pump  ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:  PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA ¿ 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA ¿ 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA ¿ 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH     1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA ¿ 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA ¿ 21-2112-0100-51 (Devices, Class I). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1662-2025&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "CADD Solis HSPCA Pump  ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:  PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA ¿ 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA ¿ 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA ¿ 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH     1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA ¿ 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA ¿ 21-2112-0100-51",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "Smiths Medical ASD, Inc."
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1662-2025/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-1662-2025: Smiths Medical ASD, Inc. is recalling CADD Solis HSPCA Pump  ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:  PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA ¿ 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA ¿ 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA ¿ 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH     1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA ¿ 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA ¿ 21-2112-0100-51",
      "text": "The FDA logged enforcement record Z-1662-2025: Smiths Medical ASD, Inc. is voluntarily recalling CADD Solis HSPCA Pump  ***Updated 5/27/26*** Scope updated to apply specifically to customers using the wireless system. Affected products associated with the change in scope:  PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, ENGLISH, 1/EA 21-2111-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, TRIAL, LOANER, ENGLISH, 1/EA 21-2111-0403-78 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, ENGLISH, 1/EA 21-2112-0403-51 PUMP KIT, CADD SOLIS HPCA, MDL 2110, V4.3, YELLOW, FRENCH, 1/EA 21-2112-0403-02 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2111-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA ¿ 21-2111-0300-50 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA ¿ 21-2112-0300-01 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, FRENCH 1/EA ¿ 21-2112-0300-02 PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA 21-2112-0300-50 PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA ¿ 21-2101-51 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA 21-2111-0401-78 PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, ENGLISH 1/EA 21-2111-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA ¿ 21-2112-0400-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA ¿ 21-2111-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0401-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA ¿ 21-2111-0402-51 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH     1/EA 21-2112-0402-78 PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA ¿ 21-2112-0402-51 PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA ¿ 21-2111-0100-51 PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA ¿ 21-2112-0100-51 — reason: Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.. Status: Ongoing — the recall is still being carried out.. Initiated 2025-04-10. Affected lots/codes: Affected CADD-Solis Software Versions: 4.0, 4.1, 4.2, 4.2.1, 4.3  UDI-DI:  4546027095675 4546027095682 10610586034787 10610586038778 10610586038785 10610586038808 10610586038822 10610586039232 10610586041037 10610586041051 15019517099690 15019517101003 15019517101010 15019517101034 15019517101041 15019517128697 15019517135343 15019517154290 15019517154887 15019517154924 15019517154979 15019517154986 15019517154993 15019517155006 15019517155013 15019517155044 15019517246735 15019517247008 15019517247169 15019517247176 15019517247183 15019517247206 00610586038900 00610586039310 04546027095620 04546027095637 04546027095644 04546027095668 04546027095699 04546027095705 04546027095712 04546027095729 04546027095736 04546027095897 04546027095903 04546027095910 04546027422983 04546027424116 04546027424130 04546027424147 04546027424178 04546027424185 04546027424192 04546027424208 04546027424598 05019517154934 05019517155108 10610586031854 10610586032950 10610586033933 10610586033940 10610586035357 10610586035395 10610586035401 10610586035418 10610586035432 10610586035456 10610586035494 10610586035548 10610586038792 10610586038839 10610586038860 10610586038891 10610586038914 10610586038921 10610586038938 10610586038952 10610586038969 10610586038983 10610586038990 10610586039010 10610586039041 10610586039065 10610586039072 10610586039089 10610586039195 10610586039256 10610586039270 10610586039409 10610586039416 10610586039430 10610586039447 10610586039454 10610586039461 10610586039478 10610586039485 10610586039492 10610586041822 30610586035337 30610586035344 30610586035368 30610586035511 30610586039205 30610586039397 35019517128677 35019517128684 35019517128691 35019517128721 35019517128738 35019517128745 35019517128769 35019517128783 35019517128837 35019517135316 35019517135323 35019517135330 35019517135354 35019517135361 35019517154850 35019517154867 35019517154874 35019517154904 35019517154911 35019517154942 35019517154959 35019517154966 35019517155055 35019517155062 35019517155079 35019517155086 35019517155130. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1662-2025&lang=eng",
        "value": "Devices recall, status Ongoing",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1662-2025&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1662-2025/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-1662-2025 as Class I",
      "text": "The FDA assigned this recall Class I: Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1662-2025&lang=eng",
        "value": "Class I",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1662-2025&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1662-2025/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-1662-2025",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-1662-2025",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class I",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Ongoing — the recall is still being carried out.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "Smiths Medical ASD, Inc.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "Pumps may experience Wireless Connection Modules intermittent connection alarms, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2025-04-10",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "Affected CADD-Solis Software Versions: 4.0, 4.1, 4.2, 4.2.1, 4.3  UDI-DI:  4546027095675 4546027095682 10610586034787 10610586038778 10610586038785 10610586038808 10610586038822 10610586039232 10610586041037 10610586041051 15019517099690 15019517101003 15019517101010 15019517101034 15019517101041 15019517128697 15019517135343 15019517154290 15019517154887 15019517154924 15019517154979 15019517154986 15019517154993 15019517155006 15019517155013 15019517155044 15019517246735 15019517247008 15019517247169 15019517247176 15019517247183 15019517247206 00610586038900 00610586039310 04546027095620 04546027095637 04546027095644 04546027095668 04546027095699 04546027095705 04546027095712 04546027095729 04546027095736 04546027095897 04546027095903 04546027095910 04546027422983 04546027424116 04546027424130 04546027424147 04546027424178 04546027424185 04546027424192 04546027424208 04546027424598 05019517154934 05019517155108 10610586031854 10610586032950 10610586033933 10610586033940 10610586035357 10610586035395 10610586035401 10610586035418 10610586035432 10610586035456 10610586035494 10610586035548 10610586038792 10610586038839 10610586038860 10610586038891 10610586038914 10610586038921 10610586038938 10610586038952 10610586038969 10610586038983 10610586038990 10610586039010 10610586039041 10610586039065 10610586039072 10610586039089 10610586039195 10610586039256 10610586039270 10610586039409 10610586039416 10610586039430 10610586039447 10610586039454 10610586039461 10610586039478 10610586039485 10610586039492 10610586041822 30610586035337 30610586035344 30610586035368 30610586035511 30610586039205 30610586039397 35019517128677 35019517128684 35019517128691 35019517128721 35019517128738 35019517128745 35019517128769 35019517128783 35019517128837 35019517135316 35019517135323 35019517135330 35019517135354 35019517135361 35019517154850 35019517154867 35019517154874 35019517154904 35019517154911 35019517154942 35019517154959 35019517154966 35019517155055 35019517155062 35019517155079 35019517155086 35019517155130",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "Worldwide - US Nationwide distribution.",
            "Status": "recorded"
          }
        }
      ]
    }
  ]
}