Drug & Device Safety Atlas
Class II FDA device recall · Z-1709-2026

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Medline kits containing Olympus biopsy valves: 1. BRONCHOSCOPY, DYNJ900898I DYNJ901922G 2. FLEXIBLE BRONCHOSCOPY KIT, DYKM1417I DYKM1417J 3. LARYNGOSCOPY/BRONCHOSCOPY PACK, DYNJ58146A 4. RIGID KIT, DYNDA2138A

Why it was recalled

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
1. DYNJ900898I UDI-DI 10198459273513 (EA) 40198459273514 (CS) LOTS 25DME064 25HMD841 25JMD489 25KME329 25KMH848 25LMC029 25LMH235 25LMK305  DYNJ901922G UDI-DI 10195327640422 (EA) 40195327640423 (CS) LOTS 24KBP812  2. DYKM1417I UDI-DI 10195327456757 (EA) 40195327456758 (CS) LOTS 24DLA405 24ELA912 24GLA871 24ILA151 24JLA326  DYKM1417J UDI-DI 10198459191640 (EA) 40198459191641 (CS) LOTS 25BLA761 25BLA886 25DLA353 25ELA056 25FLA429 25HLA196 25HLA242 25KLA388 25KLA610  3. DYNJ58146A UDI-DI 10193489611977 (EA) 40193489611978 (CS) LOTS 23KDA851 24ADC096 24CDB107 24DDB887 24IDA855 24JDB494 24LDA691 25BMF825 25DMD969 25FMC760 25GMD422  4. DYNDA2138A UDI-DI 10193489874594 (EA) 40193489874595 (CS) 24DBM325 24DBM383 25DBK159

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX.

Recall record

Recalling firm
Medline Industries, LP
Product type
Devices
Recall number
Z-1709-2026
Voluntary / mandated
Voluntary: Firm initiated
Origin
Northfield, IL
Recall initiated
2026-02-16

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1709-2026 · Source: openFDA enforcement.