Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.
NAMIC Angiographic Control Syringe, labeled as follows: 1. SYRINGE,7ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70083007; 2. SYRINGE,8ML,TR/FR,RA,W/RES,-,P/G, Medline SKU 70084007; 3. MTS,SYRINGE,8ML,TR/FR,RA,NO,RES,-,P/G, Medline SKU 70084017; 4. SYRINGE,7,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80083007; 5. SYRINGE,8,ML,TR/FR,RA,W/RES,OEM,PG, Medline SKU 80084007; 6. MTO,SYRINGE,8,ML,TR/FR,RA,NO,RES,OEM,PG, Medline SKU 80084017.
Why it was recalled
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
1. Medline SKU 70083007, UDI/DI each 10193489065855, UDI/DI case 30193489065859, Lot Numbers: 0000077029, 0000081537, 0000084335, 0000087347, 0000089563, 0000113792, 0000115753, 0000122956, 0000127785, 0000131986, 0000158748, 0000175535, 0000187926; 2. Medline SKU 70084007, UDI/DI each 10193489065893, UDI/DI case 30193489065897, Lot numbers: 0000078186, 0000081539, 0000084274, 0000086756, 0000089554, 0000094842, 0000096918, 0000105613, 0000109479, 0000111244, 0000114562, 0000115421, 0000117317, 0000118251, 0000120265, 0000120769, 0000121714, 0000123466, 0000124589, 0000131984, 0000131989, 0000138071, 0000140322, 0000142850, 0000147002, 0000152069, 0000162694, 0000173507, 0000174662, 0000174773, 0000176555, 0000178806, 0000189895; 3. Medline SKU 70084017, UDI/DI each 10193489065916, UDI/DI case 20193489065913, Lot Numbers: 0000084509, 0000086823, 0000095687, 0000165551, 0000165868, 0000166300; 4. Medline SKU 80083007, UDI/DI each 10193489066753, UDI/DI case 30193489066757, Lot Numbers: 0000097388, 0000127072, 0000135012, 0000137004; 5. Medline SKU 80084007, UDI/DI each 10193489066791, UDI/DI case 30193489066795, Lot Numbers: 0000080553, 0000080919, 0000082969, 0000085041, 0000088044, 0000094514, 0000099870, 0000101882, 0000102372, 0000105155, 0000107328, 0000109442, 0000112810, 0000112966, 0000114317, 0000115744, 0000117572, 0000118722, 0000118962, 0000119950, 0000120931, 0000121305, 0000121393, 0000122583, 0000122606, 0000123102, 0000126198, 0000126566, 0000127132, 0000130798, 0000131678, 0000132838, 0000137176, 0000139543, 0000140507, 0000142057, 0000142279, 0000143110, 0000143264, 0000143714, 0000148872, 0000150059, 0000150488, 0000151054, 0000151295, 0000151300, 0000151768, 0000152754, 0000153438, 0000154148, 0000156028, 0000158045, 0000158122, 0000159113, 0000159652, 0000160479, 0000160753, 0000161269, 0000162083, 0000162090, 0000162336, 0000162895, 0000164540, 0000164541, 0000166504, 0000166983, 0000167995, 0000168000, 0000169147, 0000169560, 0000169793, 0000170823, 0000171778, 0000173402, 0000174087, 0000174697, 0000174747, 0000176011, 0000176240, 0000176850, 0000177923, 0000179193, 0000179414, 0000180511, 0000182186, 0000182779, 0000182782, 0000185643, 0000186234, 0000186874, 0000187362, 0000187363, 0000188837, 0000188842, 0000188845; 6. Medline SKU 80084017, UDI/DI each 10193489066814, UDI/DI case 20193489066811, Lot Numbers: 0000101535, 0000103862, 0000105095, 0000106954, 0000108237, 0000122887, 0000126590, 0000127159, 0000128360, 0000128852, 0000132717, 0000134047, 0000135616, 0000140813, 0000144899, 0000148530, 0000149322, 0000152430, 0000161279, 0000164538, 0000169107, 0000173193, 0000173256, 0000182141, 0000183619, 0000187345, 0000188399.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Recall record
- Recalling firm
- Medline Industries, LP
- Product type
- Devices
- Recall number
- Z-1713-2026
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Origin
- Northfield, IL
- Recall initiated
- 2026-02-27
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1713-2026 · Source: openFDA enforcement.