  "@context": {
    "schema": "https://schema.org/",
    "au": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native#",
    "@vocab": "https://schema.org/",
    "claims": {
      "@id": "au:claims",
      "@container": "@list"
    },
    "epistemicStatus": "au:epistemicStatus",
    "occurred": "au:occurred",
    "supported": "au:supported",
    "adjudged": "au:adjudged",
    "operative": "au:operative",
    "grounding": "au:grounding",
    "groundingType": "au:groundingType",
    "adjudicator": "au:adjudicator",
    "source": "au:source",
    "sourceUrl": "au:sourceUrl",
    "value": "au:value",
    "claimType": "au:claimType",
    "reviewStatus": "au:reviewStatus",
    "commercialInfluence": "au:commercialInfluence",
    "falsifier": "au:falsifier"
  },
  "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1746-2025.jsonld",
  "@type": "schema:Dataset",
  "name": "Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35",
  "description": "FDA recall Z-1746-2025 — Ethicon Endo-Surgery Inc is recalling Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35 (Devices, Class I). The enforcement record is RECORDED; the classification is a DETERMINATION (FDA's regulatory verdict); the boundary is INTERPRETATION (a recall ≠ harm to a specific person). Source: openFDA. CC0.",
  "license": "https://creativecommons.org/publicdomain/zero/1.0/",
  "isAccessibleForFree": true,
  "isPartOf": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
  "rdfs:isDefinedBy": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/vocab/agent-native",
  "seeAlso": "https://agenticu.ryanjhunter.workers.dev/commons/the-research-object",
  "dateModified": "2026-06-24T07:02:03.626072+00:00",
  "isBasedOn": {
    "@type": "schema:Dataset",
    "name": "openFDA drug & device enforcement (recall) reports",
    "url": "https://open.fda.gov/apis/drug/enforcement/",
    "creator": {
      "@type": "schema:GovernmentOrganization",
      "name": "U.S. Food & Drug Administration"
    }
  },
  "sameAs": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1746-2025&lang=eng",
  "about": {
    "@type": "schema:MedicalDevice",
    "name": "Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35",
    "manufacturer": {
      "@type": "schema:Organization",
      "name": "Ethicon Endo-Surgery Inc"
    }
  },
  "claims": [
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1746-2025/enforcement-record",
      "@type": "schema:Claim",
      "name": "FDA recall Z-1746-2025: Ethicon Endo-Surgery Inc is recalling Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35",
      "text": "The FDA logged enforcement record Z-1746-2025: Ethicon Endo-Surgery Inc is voluntarily recalling Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35 — reason: Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.. Status: Ongoing — the recall is still being carried out.. Initiated 2025-04-22. Affected lots/codes: UDI-DI GTIN 10705036014591  Packaging Batches: 917C65 895C93 918C15 938C92 936C69 929C38 948C78 990C81 979C81 400D82 962C74 506D29 379D80 413D56 418D41 419D47 415D43 427D07 434D35 442D78 442D83 444D96 456D87 455D03 493D27 468D88 482D30 502D90 489D62 134D80 112D03 155D33 159D49 181D11 194D77 125D04 167D13 223D59 175D72 234D11 238D70 261D10 270D69 317D74 194D76 349D54 347D73 317D73 356D54 326D66 319D76 338D99 336D39 356D53 367D26 378D04 214D17. A RECORDED FDA enforcement action (occurred=true).",
      "claimType": "official_fact",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": true,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "SourceCitation",
        "source": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1746-2025&lang=eng",
        "value": "Devices recall, status Ongoing",
        "adjudicator": {
          "name": "openFDA drug & device enforcement (recall) reports — U.S. FDA",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1746-2025&lang=eng"
        }
      },
      "falsifier": "A RECORD of measured/reported events (occurred=true) from openFDA drug & device enforcement (recall) reports — U.S. FDA; superseded when openFDA drug & device enforcement (recall) reports — U.S. FDA publishes its next release."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1746-2025/fda-classification",
      "@type": "schema:Claim",
      "name": "FDA classified recall Z-1746-2025 as Class I",
      "text": "The FDA assigned this recall Class I: Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.. This is the FDA's official risk classification — its regulatory DETERMINATION of the hazard category (a verdict constituted by the FDA's recognition, occurred=FALSE), adjudged=TRUE because the FDA issued it. It is NOT a measurement of harm to any specific person.",
      "claimType": "regulatory_determination",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": true,
        "operative": true
      },
      "grounding": {
        "groundingType": "AdjudicativeAct",
        "source": "U.S. FDA recall classification (openFDA enforcement)",
        "sourceUrl": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1746-2025&lang=eng",
        "value": "Class I",
        "adjudicator": {
          "name": "U.S. FDA recall classification (openFDA enforcement)",
          "url": "https://www.accessdata.fda.gov/scripts/enforcement/enforce_rpt-Product-Tabs.cfm?action=select&recall_number=Z-1746-2025&lang=eng"
        }
      },
      "falsifier": "occurred=FALSE: a regulatory DETERMINATION (an authority's classification/judgment) has NO material truthmaker independent of recognition — it is constituted by U.S. FDA recall classification (openFDA enforcement)'s say-so (a Searle institutional fact), not measured. adjudged=TRUE because U.S. FDA recall classification (openFDA enforcement) (an external authority) issued it. Distinct from a recorded measurement (which has an independent truthmaker) and from a modeled figure (a computed quantity, not a verdict). Challengeable; superseded if the authority reclassifies."
    },
    {
      "@id": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/recalls/Z-1746-2025/boundary",
      "@type": "schema:Claim",
      "name": "What an FDA recall is — and is NOT",
      "text": "A recall is a RECORDED FDA enforcement action — the FDA logged that a firm is removing or correcting a product, classed by the FDA's risk judgment. It does NOT establish that any specific unit is affected or caused harm; causation is not established. This record informs and compares — it is NOT medical advice and does not tell anyone to start or stop a drug or device. For a medical decision, route to the FDA recall record and a licensed pharmacist or prescriber.",
      "claimType": "interpretation",
      "reviewStatus": "current",
      "commercialInfluence": false,
      "epistemicStatus": {
        "occurred": false,
        "supported": true,
        "adjudged": false,
        "operative": true
      },
      "grounding": {
        "groundingType": "Interpretation",
        "source": "Atlas Commons L5 boundary (enforcement-action ≠ harm causation)",
        "sourceUrl": "https://drug-device-safety-atlas.ryanjhunter.workers.dev/",
        "value": "a recorded enforcement action, not proof of harm or causation; not medical advice",
        "adjudicator": null
      },
      "falsifier": "occurred=FALSE (a framing, not an event) and adjudged=FALSE (an in-house reading, not an authority's verdict) — the LOWEST-warrant claim here. A flat representation would render this identically to the official facts; the four-status does not."
    }
  ],
  "hasPart": [
    {
      "@type": "schema:ItemList",
      "name": "FDA recall details — Z-1746-2025",
      "numberOfItems": 9,
      "itemListElement": [
        {
          "@type": "schema:ListItem",
          "position": 1,
          "item": {
            "Field": "Recall number",
            "Value": "Z-1746-2025",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 2,
          "item": {
            "Field": "Product type",
            "Value": "Devices",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 3,
          "item": {
            "Field": "Classification",
            "Value": "Class I",
            "Status": "determination"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 4,
          "item": {
            "Field": "Status",
            "Value": "Ongoing — the recall is still being carried out.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 5,
          "item": {
            "Field": "Recalling firm",
            "Value": "Ethicon Endo-Surgery Inc",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 6,
          "item": {
            "Field": "Reason",
            "Value": "Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 7,
          "item": {
            "Field": "Initiated",
            "Value": "2025-04-22",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 8,
          "item": {
            "Field": "Lot/code info",
            "Value": "UDI-DI GTIN 10705036014591  Packaging Batches: 917C65 895C93 918C15 938C92 936C69 929C38 948C78 990C81 979C81 400D82 962C74 506D29 379D80 413D56 418D41 419D47 415D43 427D07 434D35 442D78 442D83 444D96 456D87 455D03 493D27 468D88 482D30 502D90 489D62 134D80 112D03 155D33 159D49 181D11 194D77 125D04 167D13 223D59 175D72 234D11 238D70 261D10 270D69 317D74 194D76 349D54 347D73 317D73 356D54 326D66 319D76 338D99 336D39 356D53 367D26 378D04 214D17",
            "Status": "recorded"
          }
        },
        {
          "@type": "schema:ListItem",
          "position": 9,
          "item": {
            "Field": "Distribution pattern",
            "Value": "US distribution nationwide. OUS distribution to EMEA, LATAM, APAC, and Canada.",