Drug & Device Safety Atlas
Class II FDA device recall · Z-1789-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

RKNXC0003, RHEO KNEE XC 3 YR WARRANTY, MD, CE, External lower limb prosthetic component

Why it was recalled

Due to firmware issues with the prosthetic knee, there is the potential for unintended warnings and device shutdown which could result in patient falls.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI: 05690977467804/ Serial Numbers: HF366771 HF366854 HF366819 HF363224 HF366748 HF364570 HF366805 HF364686 HF366831 HF365048 HF366880 HF365049 HF366757 HF365311 HF366787 HF365378 HF366815 HF365447 HF366826 HF365546 HF366838 HF365787 HF366874 HF365810 HF362144 HF366081 HF366755 HF366229 HF366759 HF366409 HF366778 HF366647 HF366795 HF366648 HF366811 HF366650 HF366817 HF366651 HF366823 HF366652 HF366828 HF366653 HF366836 HF366659 HF366849 HF366661 HF366860 HF366665 HF366878 HF366672 HF366893 HF366673 HF366738 HF366678 HF366754 HF366680 HF366756 HF366681 HF366758 HF366690 HF366762 HF366692 HF366775 HF366693 HF366782 HF366696 HF366793 HF366697 HF366802 HF366698 HF366806 HF366699 HF366812 HF366700 HF366816 HF366701 HF366818 HF366706 HF366821 HF366707 HF366824 HF366708 HF366827 HF366714 HF366829 HF366717 HF366834 HF366718 HF366837 HF366719 HF366839 HF366720 HF366853 HF366722 HF366859 HF366723 HF366862 HF366724 HF366875 HF366732 HF366879 HF366733 HF366884 HF366734 HF366895 HF366735 HF366737

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.

Recall record

Recalling firm
Ossur H / F
Product type
Devices
Recall number
Z-1789-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Reykjavik, N/A
Recall initiated
2024-03-25

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1789-2024 · Source: openFDA enforcement.