Drug & Device Safety Atlas
Class I FDA device recall · Z-1809-2025

Class I — the FDA judged there is a reasonable probability that use of the product would cause serious harm or death.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone.

Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A

Why it was recalled

Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
1. CRANI ADD A PACK, SKU DYNJ30926O, UDI-DI 10195327539153 (EA), 40195327539154 (CASE)  KIT LOTS 24AMF811   2. CRANIOTOMY, SKU CDS983611N, UDI-DI 10195327466626 (EA), 40195327466627 (CASE)  KIT LOTS   23KBA720 23LBI611 23LBJ586  3. CRANIOTOMY CDS, SKU CDS983467J, UDI-DI 10195327561116 (EA),  40195327561117 (CASE)  KIT LOTS  24ADA818 24ADC216 24BDA929 24DDB447 24EDA275 24EDA751 24GDB213 24IDA327 24IDB043   4. CRANIOTOMY PACK, SKU DYNJ85807A, UDI-DI 10195327637392 (EA), 40195327637393 (CASE)  KIT LOTS 24CBB843 24EBC889 24GBY029 ;   SKU DYNJ85927, UDI-DI 10195327575571 (EA), 40195327575572 (CASE)  KIT LOTS 24BMD310 24EMI836   5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A UDI-DI 10195327635466 (EA), 40195327635467 (CASE)  KIT LOTS 24CMA496 24HMI081 24HMI978

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.

Recall record

Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product type
Devices
Recall number
Z-1809-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Northfield, IL
Recall initiated
2025-04-22

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1809-2025 · Source: openFDA enforcement.