Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Panorama HFO- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781350
Why it was recalled
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
UDI-DI: (1) 782152 N/A (2) 782110 (01)00884838099364(21) Serial Numbers: Serial Number 37019 37067 37149 37188 37184 37034 37036 37058 37167 37173 37183 37195 37197 37009 37011 37013 37020 37027 37031 37050 37055 37080 37093 37100 37121 37127 37132 37137 37141 37143 37150 37157 37170 37180 37193 37194 37023 37044 37066 37071 37106 37124 37160 37175 37182 37192 37022 37068 37082 37129 37134 37145 37156 37171 37181 37187 19083 37161 37178 37088 37024 37098 37135 37139 37176 37153 37060 37116 37196 37015 37073 37109 37122 37133 37162 37185 37026 37096 37177 37002 37012 37017 37029 37039 37041 37047 37048 37052 37053 37056 37077 37083 37084 37086 37087 37095 37102 37104 37107 37110 37113 37120 37123 37125 37128 37130 37136 37142 37158 37159 37163 37165 37168 37189 37069 37174 37126
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Recall record
- Recalling firm
- Philips North America Llc
- Product type
- Devices
- Recall number
- Z-1855-2024
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- Letter
- Origin
- Cambridge, MA
- Recall initiated
- 2024-05-03
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1855-2024 · Source: openFDA enforcement.