Class II FDA device recall · Z-1856-2024

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Marlin 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781474

Why it was recalled

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI: N/A  Serial Numbers:  Serial Number 79113 79118 79127 79157 79122 79147 79105 79150 79152 79183 79192 79102 79108 79112 79115 79116 79153 79156 79162 79163 79164 79168 79188 79189 79100 79126 79179 79143 79159 79104 79132 79160 79182 79165 79107 79158 79175 79136 79138 79139 79184 79176 79180 79171 79172 79111 79125 79134 79135 79148 79177 79181 79173 79187 79190 79133 79110 79101 79117 79166 79124 79123 79130 79174 79178 79151 79114 79103 79106 79121 79128 79129 79137 79140 79155 79161 79167 79169 79170 79185

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Recall record

Recalling firm
Philips North America Llc
Product type
Devices
Recall number
Z-1856-2024
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Cambridge, MA
Recall initiated
2024-05-03

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1856-2024 · Source: openFDA enforcement.