Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Dexcom G7 iOS Application, Ref: SW12300 used in conjunction with the Dexcom G7 Continuous Glucose Monitoring (CGM) system
Why it was recalled
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose alerts will not trigger, there is potential for the missed detection of a hyperglycemic or hypoglycemic event or a treatment decision made based off incorrect data. There is no impact to transmitter communication with concurrently connected displays. A concurrently connected receiver, Apple Watch, and/or insulin pump will continue to receive EGVs directly from the transmitter when the app is in the problematic state. There is no impact to Automated Insulin Dosing (AID) therapy as the AID system continues to receive EGVs directly from the transmitter.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
Software version 2.7.0/ UDI: 00386270001863
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide Distribution: US (nationwide) and OUS (foreign) to countries of: United Arab Emirates (AE), Austria (AT), Australia (AU), Belgium (BE), Bahrain (BH), Canada (CA), Switzerland (CH), Czech Republic (CZ), Germany (DE), Denmark (DK), Spain (ES), Finland (FI), France (FR), United Kingdom (GB), Hong Kong (HK), Ireland (IE), Israel (IL), Iceland (IS), Italy (IT), Jordan (JO), Korea, Repulic (KR), Kuwait (KW), Lebanon (LB), Lithuania (LT), Luxemburg (LU), Malta (MT), Netherlands (NL), Norway (NO), New Zealand (NZ), Oman (OM), Poland (PL), Portugal (PT), Qatar (QA), Saudi Arabia (SA), Sweden (SE), Slovenia (SI), Slovakia (SK), Turkey (TR) and South Africa (ZA).
Recall record
- Recalling firm
- Dexcom, Inc.
- Product type
- Devices
- Recall number
- Z-1906-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Initial notification
- N/A
- Origin
- San Diego, CA
- Recall initiated
- 2025-05-05
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.
To report a problem you experienced, use
FDA MedWatch.
Verify this record at FDA recalls ·
Recall Z-1906-2025 · Source: openFDA enforcement.