Class II FDA device recall · Z-1970-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium Closed Male Luer With Priming Cap, REF: 22000-B007T; Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, REF: 22600-0007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22603-B007T; Bonded Texium Closed Male Luer with priming cap Back Check Valve 2 SmartSite Y-Sites, REF: 24010-0007T; SmartSite Bag Access Non-Vented 0.2 MicronFilter Bonded Texium Closed Low Sorbing Tubing (PE Lined) Male Luer with priming cap SmartSite Y-site, REF: 24301-0007T; SmartSite Bag Access Non-Vented Low Sorbing Texium Back Check Valve 2 SmartSite Y-Sites, REF: 24601-B007T. BD 31-IN SEC SET 20DP W/BAG ACCESS PORT TEXIUMTM AND HANGER LOW SORBING, REF: 70001B-07T; BD 20-IN EXTENSION SET W/0.2 MICRON FILTER ONE SmartSite VALVE AND Texium, REF: 20350ET; BD Secondary Set 20 Drop Vented Non Vented With Texium & Hanger, REF: 40000-07T; BD 20 DROP SECONDARY SET SmartSite VALVE BAG ACCESS W/TEXIUM CLOSED MALE LUER HANGER, REF: 10013364T; BD 32-IN GRAV SET MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, REF: 4030B-07T; BD Texium Smartsite MICROBORE EXTENSION SET 60 IN, REF: 22005E-07T

Why it was recalled

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
REF/UDI-DI/Lot(Expiration): 10013361T/10885403223204/23025303(February 10, 2026), 23035235(March 11, 2026), 23045223(April 6, 2026), 23075395(July 17, 2026), 23085227(August 8, 2026), 23115150(November 8, 2026), 23115151(November 8, 2026), 23125167(December 5, 2026), 24025684(February 6, 2027), 24045377(April 23, 2027), 24055576(May 27, 2027), 24075224(July 26, 2027), 24105079(October 11, 2027), 24115159(November 7, 2027);  10321213T/10885403228971/23035169(March 15, 2026), 23045174(April 4, 2026), 23085140(August 3, 2026), 23095142(September 11, 2026), 23105297(October 12, 2026), 24025686(February 6, 2027);  22000-B007T/10885403245022/23025320(February 13, 2026), 23045224(April 11, 2026), 23065209(June 10, 2026), 23075394(July 21, 2026), 23095143(September 11, 2026), 23105133(October 6, 2026), 24015278(January 13, 2027), 24025685(February 6, 2027), 24036209(March 25, 2027), 24075260(July 26, 2027), 24066653(June 24, 2027);  22600-0007T/10885403239595/23095151(September 11, 2026), 24036205(March 25, 2027), 24055601(May 27, 2027);  22602-B007T/10885403240935/23045501(April 20, 2026), 23095152(September 11, 2026), 24015186(January 9, 2027), 24026152(February 27, 2027), 24036206(March 25, 2027), 24045384(April 23, 2027), 24115032(November 2, 2027);  22603-B007T/10885403240942/23035482(March 23, 2026), 23035504(March 25, 2026), 23055379(May 23, 2026), 23065210(June 12, 2026), 23075056(July 13, 2026), 23085141(August 3, 2026), 23095808(September 14, 2026), 23105296(October 12, 2026), 23125107(December 4, 2026), 24015185(January 9, 2027), 24025681(February 6, 2027), 24026153(February 27, 2027), 24026154(February 27, 2027), 24036207(March 25, 2027), 24045381(April 23, 2027), 24055577(May 27, 2027), 24066654(June 24, 2027), 24085243(August 26, 2027), 24095223(September 23, 2027), 24115312(December 2, 2027);  24010-0007T/10885403271021/23025108(February 8, 2026), 23035249(March 14, 2026), 23075392(July 18, 2026), 23085223(August 8, 2026), 23095763(September 19, 2026), 23105136(October 6, 2026), 23115054(November 3, 2026), 24015279(January 13, 2027), 24015386(January 17, 2027), 24025683(February 6, 2027), 24026149(February 27, 2027), 24045382(April 23, 2027), 24055598(May 27, 2027), 24075225(July 26, 2027), 24085240(August 26, 2027), 24095220(September 23, 2027), 24115311(November 14, 2027), 24066656(June 24, 2027);  24301-0007T/10885403223198/23025183(February 9, 2026), 23025238(February 10, 2026), 23035338(March 16, 2026), 23045222(April 12, 2026), 23075393(July 20, 2026), 23085226(August 8, 2026), 23095764(September 19, 2026), 23105284(October 11, 2026), 23115055(November 3, 2026), 24026089(February 21, 2027), 24026150(February 27, 2027), 24036178(March 25, 2027), 24045383(April 23, 2027), 24055599(May 27, 2027), 24075212(July 26, 2027), 24085241(August 26, 2027), 24095221(September 23, 2027), 24115279(November 12, 2027), 24015474(January 20, 2027);  24601-B007T/10885403240959/23035170(March 9, 2026), 23055380(May 24, 2026), 23065280(June 13, 2026), 23085142(August 3, 2026), 23095153(September 11, 2026), 23105285(October 11, 2026), 23115052(November 3, 2026), 24025682(February 6, 2027), 24026151(February 27, 2027), 24036208(March 25, 2027), 24055600(May 27, 2027), 24075261(July 26, 2027), 24085242(August 26, 2027), 24105106(October 22, 2027), 24066655(June 24, 2027), 24015475(January 20, 2027);  70001B-07T/10885403240997/24109019(October 2, 2027), 24109018(October 2, 2027), 24089215(August 9, 2027), 24089214(August 9, 2027), 24019266(January 16, 2027), 23109293(October 23, 2026), 24099179(September 5, 2027), 24049266(April 15, 2027), 24039248(March 14, 2027);  20350ET/10885403229756/24129026(November 13, 2027), 24079037(July 2, 2027), 24059361(May 15, 2027), 25029088(February 4, 2028), 24049028(April 2, 2027);  40000-07T/10885403225628/24119116(November 6, 2027), 24089019(August 1, 2027), 24079173(July 3, 2027), 24059480(May 28, 2027), 24059152(May 7, 2027), 24019267(January 16, 2027), 23119354(November 29, 2026), 24049267(April 15, 2027), 24039212(March 12, 2027), 23109275(October 19, 2026), 23079215(July 20, 2026);  10013364T/10885403198915/24109021(October 2, 2027), 24109020(October 2, 2027), 24099172(September 5, 2027), 24089059(August 5, 2027), 24089058(August 4, 2027), 24079175(July 3, 2027), 24079174(July 3, 2027), 24059506(May 30, 2027), 24059469(May 28, 2027), 24059151(May 7, 2027), 24039059(March 4, 2027), 24019435(January 31, 2027), 24019268(January 16, 2027), 24019433(January 30, 2027), 24019269(January 16, 2027), 23119087(November 6, 2026), 23129229(December 13, 2026), 23129230(December 13, 2026), 23059178(May 24, 2026), 23059177(May 23, 2026), 23059156(May 16, 2026), 23049061(April 26, 2026), 24059150(May 7, 2027), 24049265(April 15, 2027), 23109175(October 13, 2026), 23109006(October 2, 2026), 23099219(September 12, 2026), 23099217(September 11, 2026), 23059179(May 29, 2026);  4030B-07T/10885403240867/24089216(August 9, 2027), 24059484(May 28, 2027), 24039060(March 4, 2027), 25029393(February 27, 2028), 23109296(October 23, 2026);  22005E-07T/10885403250781/23029335(February 24, 2026)

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Recall record

Recalling firm
BD SWITZERLAND SARL
Product type
Devices
Recall number
Z-1970-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Eysins
Recall initiated
2025-05-12

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1970-2025 · Source: openFDA enforcement.