Drug & Device Safety Atlas
Class II FDA device recall · Z-1971-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060

Why it was recalled

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
UDI-DI/Lot(Expiration): MY8003/10885403221651/23125144(December 5, 2026), 24015468(January 20, 2027), 24025611(February 5, 2027), 24026390(February 27, 2027), 24036107(March 25, 2027), 24045647(April 24, 2027), 24055859(May 27, 2027), 24095513(September 24, 2027), 92217803(June 27, 2025), 92225801(September 15, 2025), 92231303(November 9, 2025), 92300201(January 2, 2026), 92308702(March 28, 2026), 92309501(April 5, 2026), 92313601(May 16, 2026), 92315701(June 6, 2026)  MY8005/10885403221668/23125145(December 5, 2026), 24015467(January 20, 2027), 24026387(February 27, 2027), 24045646(April 24, 2027), 24055858(May 27, 2027), 24075164(July 26, 2027), 24095512(September 24, 2027), 92215201(June 1, 2025), 92220802(July 27, 2025), 92225502(September 12, 2025), 92315102(May 31, 2026);  MY8010/10885403221675/24015312(January 15, 2027), 24025610(February 5, 2027), 24045645(April 24, 2027), 24055857(May 27, 2027), 24066953(June 25, 2027), 24075165(July 26, 2027), 24085486(August 26, 2027), 24095511(September 24, 2027), 24105167(October 25, 2027), 92218601(July 5, 2025), 92222902(August 17, 2025), 92223601(August 24, 2025), 92229201(October 19, 2025), 92230701(November 3, 2025), 92231301(November 9, 2025), 92231302(November 9, 2025), 92307402(March 15, 2026), 92311601(April 26, 2026), 92315101(May 31, 2026);  MY8020/10885403221682/23115069(November 3, 2026), 24015311(January 15, 2027), 24025935(February 7, 2027), 24035566(March 5, 2027), 24036106(March 25, 2027), 24045644(April 24, 2027), 24055856(May 27, 2027), 24066952(June 25, 2027), 24075163(July 26, 2027), 24085485(August 26, 2027), 24095083(September 9, 2027), 24105196(October 30, 2027), 92216401(June 13, 2025), 92217801(June 27, 2025), 92218001(June 29, 2025), 92219202(July 11, 2025), 92231901(November 15, 2025), 92300205(January 2, 2026), 92300902(January 9, 2026), 92309503(April 5, 2026), 92310101(April 11, 2026), 92313002(May 10, 2026), 92313602(May 16, 2026)  MY8030/10885403230523/23115070(November 3, 2026), 23125143(December 5, 2026), 24015264(January 12, 2027), 24015644(January 27, 2027), 24026389(February 27, 2027), 24036105(March 25, 2027), 24045643(April 24, 2027), 24055855(May 27, 2027), 24075162(July 26, 2027), 24085484(August 26, 2027), 24095509(September 24, 2027), 24115110(November 5, 2027), 92225501(September 12, 2025), 92233602(December 2, 2025), 92307403(March 15, 2026), 92308701(March 28, 2026), 92309502(April 5, 2026), 92315702(June 6, 2026), 92316402(June 13, 2026), 24066951(June 25, 2027);  MY8060/10885403221699/23115496(November 28, 2026), 23125393(December 18, 2026), 24015102(January 8, 2027), 24026388(February 27, 2027), 24036104(March 25, 2027), 24045366(April 23, 2027), 24055854(May 27, 2027), 24075166(July 26, 2027), 24095596(September 24, 2027), 24115111(November 14, 2027), 92216402(June 13, 2025), 92222002(August 8, 2025), 92222901(August 17, 2025), 92224303(August 31, 2025), 92224301(August 31, 2025), 92226502(September 22, 2025), 92228501(October 12, 2025), 92230702(November 3, 2025), 92230703(November 3, 2025), 92230704(November 3, 2025), 92300202(January 2, 2026), 92300203(January 2, 2026), 92306703(March 8, 2026), 92306701(March 8, 2026), 92306702(March 8, 2026), 92311602(April 26, 2026), 92312302(May 3, 2026), 92313001(May 10, 2026), 92316601(June 15, 2026), 92316602(June 15, 2026)

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Recall record

Recalling firm
BD SWITZERLAND SARL
Product type
Devices
Recall number
Z-1971-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Eysins
Recall initiated
2025-05-12

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1971-2025 · Source: openFDA enforcement.