Drug & Device Safety Atlas
Class II FDA device recall · Z-1972-2025

Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.

This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.

BD Texium: 3 mL, REF: MY8003-0006; 5 mL, REF: MY8005-0006; 10 mL, REF: MY8010-0006; 20 mL, REF: MY8020-0006; 30 mL, REF: MY8030-0006; 60 mL, REF: MY8060-0006

Why it was recalled

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

Is your unit affected?

The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.

A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.

Show the FDA's listed lot / code information
REF/UDI-DI/Lot(Expiration): MY8003-0006/7613203018821/24015489(January 22, 2027), 24025879(February 7, 2027), 24045651(April 24, 2027), 24055852(May 27, 2027), 24066957(June 25, 2027), 24075170(July 26, 2027), 24095519(September 24, 2027), 92215801(June 7, 2025), 92217802(June 27, 2025), 92220702(July 26, 2025), 92221502(August 3, 2025), 92303201(February 1, 2026), 92303202(February 1, 2026), 92312301(May 3, 2026), 92316401(June 13, 2026);  MY8005-0006/7613203018838/24025878(February 7, 2027), 24026382(February 27, 2027), 24036112(March 25, 2027), 24045650(April 24, 2027), 24055851(May 27, 2027), 24066956(June 25, 2027), 24075169(July 26, 2027), 24095517(September 24, 2027), 92229203(October 19, 2025), 92229204(October 19, 2025), 92229701(October 24, 2025), 92230602(November 2, 2025), 92304101(February 10, 2026), 92308703(March 28, 2026), 92314401(May 24, 2026), 92314402(May 24, 2026);  MY8010-0006/7613203018845/24025880(February 7, 2027), 24036111(March 25, 2027), 24045652(April 24, 2027), 24066958(June 25, 2027), 24075171(July 26, 2027), 24085488(August 26, 2027), 24095518(September 24, 2027), 24105168(October 25, 2027), 92217101(June 20, 2025), 92222903(August 17, 2025), 92233202(November 28, 2025), 92233201(November 28, 2025), 92300204(January 2, 2026), 92300901(January 9, 2026), 92301901(January 19, 2026), 92301902(January 19, 2026), 92303203(February 1, 2026), 92314403(May 24, 2026);  MY8020-0006/7613203018852/24025882(February 7, 2027), 24025883(February 7, 2027), 24026385(February 27, 2027), 24036110(March 25, 2027), 24045649(April 24, 2027), 24055853(May 27, 2027), 24066954(June 25, 2027), 24075168(July 26, 2027), 24085487(August 26, 2027), 24095515(September 24, 2027), 92221501(August 3, 2025), 92221503(August 3, 2025), 92222904(August 17, 2025), 92228502(October 12, 2025), 92228701(October 14, 2025), 92228702(October 14, 2025), 92229202(October 19, 2025), 92234001(December 6, 2025), 92301701(January 17, 2026), 92301702(January 17, 2026), 92302401(January 24, 2026), 92307401(March 15, 2026), 92310902(April 19, 2026);  MY8030-0006/7613203026093/24025881(February 7, 2027), 24026384(February 27, 2027), 24036109(March 25, 2027), 24045648(April 24, 2027), 24066955(June 25, 2027), 24075167(July 26, 2027), 92215802(June 7, 2025), 92219201(July 11, 2025), 92220701(July 26, 2025), 92220801(July 27, 2025), 92225802(September 15, 2025), 92226501(September 22, 2025), 92231902(November 15, 2025);  MY8060-0006/7613203018869/23125068(December 2, 2026), 24026383(February 27, 2027), 24045653(April 24, 2027), 24066879(June 24, 2027), 24075001(July 1, 2027), 24085489(August 26, 2027), 24095516(September 24, 2027), 24115310(November 5, 2027), 92218602(July 5, 2025), 92222001(August 8, 2025), 92223602(August 24, 2025), 92224302(August 31, 2025), 92225803(September 15, 2025), 92227002(September 27, 2025), 92227001(September 27, 2025), 92227901(October 6, 2025), 92233601(December 2, 2025), 92310102(April 11, 2026), 92310901(April 19, 2026);

Recall status

Ongoing — the recall is still being carried out.

Where it was distributed

Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Recall record

Recalling firm
BD SWITZERLAND SARL
Product type
Devices
Recall number
Z-1972-2025
Voluntary / mandated
Voluntary: Firm initiated
Initial notification
Letter
Origin
Eysins
Recall initiated
2025-05-12

What to do

Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.

Also available: structured data (JSON-LD) · recall record (JSON)

What this is — and is not

A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.

To report a problem you experienced, use FDA MedWatch. Verify this record at FDA recalls · Recall Z-1972-2025 · Source: openFDA enforcement.