Class II — the FDA judged the product may cause temporary or medically reversible harm, with a remote chance of serious harm.
This is the FDA's risk judgment for this recall. It describes the potential of the product as a category — it does not establish that any specific unit harmed anyone. Most recalls are Class II or III.
Brand Name: VITROS; Product Name: VITROS XT 7600 Integrated System; Model/Catalog Number: (1) 6844461, (2) 6272222; Software Version: version 3.2 to version 3.8.3 (all available software versions); Product Description: In vitro diagnostic chemistry analyzer.
Why it was recalled
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading. The issue was identified by a customer complaint, in which a CREA Urine test was processed with an RE (Reagent Expired) code, even though the diluent pack was not expired and the use-expired-reagents setting was disabled. This can result in erroneous test results and/or delaying test results.
Is your unit affected?
The recall covers only the specific lots or serial numbers the FDA listed. Enter the lot or serial number from your product's label and we'll check it against that list.
A match means your unit may be covered — stop and call your pharmacist before doing anything. No match suggests it isn't covered, but confirm anyway: code formats vary and lists can be incomplete. Your unit not being listed does not guarantee it is safe, and being listed does not mean you have been harmed.
Show the FDA's listed lot / code information
VITROS XT 7600 Integrated System; Product Code: 6844461; UDI: 10758750012343; VITROS XT 7600 System - Certified; Product Code: 6272222; UDI: 10758750012343; Range of installed serial numbers:76000109-76002327; Total systems currently installed in the US: 1,019 Total systems currently installed outside the US: 939 Total systems installed worldwide (Product Quantity Distributed): 1,958 Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured, the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today.
Recall status
Ongoing — the recall is still being carried out.
Where it was distributed
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.
Recall record
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product type
- Devices
- Recall number
- Z-2039-2025
- Voluntary / mandated
- Voluntary: Firm initiated
- Origin
- Rochester, NY
- Recall initiated
- 2025-05-09
What to do
Check the lot/code above against your product. To confirm this record and read the FDA's full notice, see the FDA recall page. Before changing anything you take or use, ask your pharmacist or prescriber — this page does not tell you to start or stop any drug or device.
Also available: structured data (JSON-LD) · recall record (JSON)
What this is — and is not
A recall is a recorded FDA enforcement action — the FDA logged that a firm is removing or correcting a product. It does not establish that any specific unit harmed anyone, and causation is not established by a recall. The class above is the FDA's risk judgment for the product category, not a finding about your unit. This page informs and compares; it is not medical advice and never tells you to start or stop a drug or device.